Instruction Manual
  <5. EtherLCD>
5-53
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM Number  1 6
Method # Method number  1 6
Stream valve 
ON time 
Warming 
up time 
0.0 seconds 
Stream valve 
OFF time 
 0.0 Main cycle 
 2
seconds 
  SYS Method Setup Screen (Main) 
Press F3 (SYS).
11/11/22
15:15:45
SYS Method Setup 
GCM #1  /  SYS #1  /  Method #1  
> - Execution  On
 - Cycle time  60.0 s
 - End of peak detect  50.0 s
 - Tracking exec  Not executed
 - Tracking coef updating  Updating
GCM Seq Copy SYS#
F1 F2 F3 F4 F5 F6
Figure 5.67 
Example of the SYS method setup screen
F1(GCM):  Returns to GCM Method Setup screen.
F2(Seq):  Displays SYS Sequential Display screen.
F3(Temp):  Displays Oven Program Set screen.
F4(EPC):  Displays Pressure Program Set screen.
F5(Copy):  Displays Copy screen of the SYS method.
F6(Sys#): Specifi es the SYS number.
The GCM number and the method number displayed in the second line on the top will succeeds 
the setup of the GCM method screen. 
Pointing the cursor to each setup and pressing ENT display each name in the second line on the 
bottom. 
Enters the time using the tenkey.
Pressing F2(Seq) navigates to the SYS sequential screen display. 
Pressing F5 (Copy) navigates to the copy screen display for the SYS method.
Pressing F6 (SYS #) displays SYS number: in the second line on the bottom.
Pressing SET/ENT on the tenkey navigates to that SYS number.
Only the SYS number set in the GCM can be selected, “Over upper/lower limit” is displayed in the 
second line on the bottom when anything else is entered.
Pressing F1 (GCM) navigates to the GCM method selection screen.
F4 (EPC) is available only when EPC has been set.
2nd Edition : May 11, 2012-00