Instruction Manual

<5. EtherLCD>
5-52
IM 11B08A01-01E
GCM Sequential Display Screen
Press F2( Seq).
11/11/22
15:15:45
GCM Sequential Display
GCM #1 / Method #1
Time Action
-5.0 StrV on time
0.0 StrV off time
50.0 S1 Peak detect end
55.0 Pause time
Main StrV
F1 F2 F3 F4 F5 F6
Figure 5.65
Example of the GCM sequential display screen
F1(Main): Displays GCM Method (Main) screen.
F3(StrV): Displays Stream Valve Set screen.
Displays the sequence of the GCM method and the SYS method set in the GCM.
When newly displayed, the sequential display from the rst sequence is displayed.
However, SYS sequence set to No for the SYS usage set is not displayed.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM number 1 6
Method # Method number 1 6
Time
Action Time Pause time
S1 to S6 Peak detect end
StrV on time, StrV off time
Stream Valve Set Screen
Press F3 (StrV).
11/11/22
15:15:45
Stream valve Set
GCM #1 / Method #1
> - Stream valve ON time -5.0 s
- Stream valve OFF time 0.0 s
Seq
F1 F2 F3 F4 F5 F6
Figure 5.66
Example of the stream valve set screen
F1(Seq): Displays Sequential Display screen.
2nd Edition : May 11, 2012-00