5.4.5 Method Setup
  GCM method setup screen (Main)
11/11/22
15:15:45
GCM Method Setup 
GCM #1  /  Method #1  
> - Main cycle  60.0 s
 - Warming up time   5.0 s
 - Stream valve ON time  -5.0 s
 - Stream valve OFF time   0.0 s
 - Pause time  55.0 s
Main Seq Sys Copy GCM#
Method#
F1 F2 F3 F4 F5 F6
Figure 5.63  Example of the GCM method setup screen
F1(Menu):  Displays the Table Menu screen.
F2(Seq):  Displays the GCM Sequential screen.
F3(SYS):  Displays the SYS Method screen.
F4(Copy):  Displays the GCM Method Copy screen.
F5(GCM #):  Specifi es the GCM number.
F6(Method #):  Specifi es the Method number.
•  Pressing F3 (SYS) displays the SYS Method Setup screen with the smallest number among 
the SYS numbers set to GCM.
•  Pressing F5 (GCM #) displays GCM number: in the second line on the bottom. Entering the 
method number refl ects the GCM number in the second line on the top and displays the 
GCM method setup assigned to that number.
•  Pressing F6 (Method #) displays Method number: in the second line on the bottom. Entering 
the method number refl ects the method number in the second line on the top and displays 
the GCM method setup assigned to that number.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM number 1 6
Method # Method number  1 6
Main cycle   10 See right seconds  Upper Limit will depend on 
sample rate
When 20 ms: 2700 seconds
When 40 ms: 5400 seconds 
When 80 ms: 10800 seconds 
When 160 ms: 21600 seconds 
* The smallest sample rate set 
for the detector used by the SYS 
belonging to the relevant GCM.
Warming up 
time 
 5.0 9999.9 seconds 
Stream valve 
ON time 
Warming 
up time 
0.0 seconds 
Stream valve 
OFF time 
 0.0 Main cycle 
 2
seconds 
Pause time   5 Main cycle 
 2
seconds 