Instruction Manual
<5. EtherLCD>
5-49
IM 11B08A01-01E
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #): Sets a detector number.
TIP
• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *
(asterisk).
• The initial screen displays Oven number 1 and Det number 1.
• When A/O chromatogram is set to None, items for Auto gain and Whole gain value are
displayed but an asterisk is given for set value.
• The initial screen displays the smallest oven number among those set in the isothermal
ovens on the Temperature Control Set.
• The initial screen displays Oven number 1 and Det number 1.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type 0) None
1) TCD
2) FID
3) FID-MC
4) FPD
Auto gain 0) None (default)
1) Individual gain
2) Whole gain *1
*1 Only for the A/O
chromatogram
Whole gain
value
0 15 0(default)
*1 Only for the A/O
chromatogram
Auto zero 0)Automatic feed (default)
1)Auto zero value
2)Disabled
* If the value is Automatic feed,
the value is set to the zero value
automatically obtained at the
auto zero time.
* If the value is Auto zero value,
the value is set to the auto zero
value at the start of the analysis.
Auto zero
value
1000.0000 1000.0000 Former chromatogram standard
value
Auto zero
time
0 21600 If the value is 0.0 seconds, the
value is set to the zero value
automatically obtained at the
start of the analysis cycle.
* Upper Limit will use the
shortest analysis cycle for the
SYS analysis cycle that the
detector belongs to.
2nd Edition : May 11, 2012-00