Instruction Manual
<5. EtherLCD>
5-48
IM 11B08A01-01E
5.4.4 Detector Setup
11/11/22
15:15:45
Detector Setup (1/2)
11/11/22
15:15:45
Detector Setup (2/2)
Oven # 1 Det # 1 ( TCD ) Oven # 1 Det # 1 ( TCD )
> - Filtering const Off > - Auto gain setting None
- Sample rate 20 ms - Full scale value 0
- Square root calc req * * * * * - Auto zero Off
- Flame detect level * * * * *
- Auto zero value 0.0000 mV
- Sense set Equal - Auto zero time 0.0 s
- Input resistor * * * * * * * * * *
Menu Oven# Det# Menu Oven# Det#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.62
Example of the detector setup screen
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #): Sets a detector number.
TIP
• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *
(asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type None
TCD
FID
FID-MC
FPD
Filtering const 0.001 1.000
Sample rate * 20 ms
40 ms
80 ms
160 ms
Square root
calc req
The value is shown except
when the FPD is installed.
No
Yes
* The square root of the area
value
Flame detect
level
–20 20 mV The item is displayed only when
the FID, FPD or FID-MC is
installed.
Sense set No amplifi cation
Tenfold
Input resistor 1 Mohm
6 Mohm
20 Mohm
60 Mohm
200 Mohm
600 Mohm
1 Gohm
2 Gohm
*: The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00