Instruction Manual

<5. EtherLCD>
5-48
IM 11B08A01-01E
5.4.4 Detector Setup
11/11/22
15:15:45
Detector Setup (1/2)
11/11/22
15:15:45
Detector Setup (2/2)
Oven # 1 Det # 1 ( TCD ) Oven # 1 Det # 1 ( TCD )
> - Filtering const Off > - Auto gain setting None
- Sample rate 20 ms - Full scale value 0
- Square root calc req * * * * * - Auto zero Off
- Flame detect level * * * * *
- Auto zero value 0.0000 mV
- Sense set Equal - Auto zero time 0.0 s
- Input resistor * * * * * * * * * *
Menu Oven# Det# Menu Oven# Det#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.62
Example of the detector setup screen
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #): Sets a detector number.
TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
If the Detector type has been set to None on the Detector Set screen, all the options are shown as *
(asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type None
TCD
FID
FID-MC
FPD
Filtering const 0.001 1.000
Sample rate * 20 ms
40 ms
80 ms
160 ms
Square root
calc req
The value is shown except
when the FPD is installed.
No
Yes
* The square root of the area
value
Flame detect
level
–20 20 mV The item is displayed only when
the FID, FPD or FID-MC is
installed.
Sense set No ampli cation
Tenfold
Input resistor 1 Mohm
6 Mohm
20 Mohm
60 Mohm
200 Mohm
600 Mohm
1 Gohm
2 Gohm
*: The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00