Instruction Manual
  <5. EtherLCD>
5-48
IM 11B08A01-01E
5.4.4 Detector Setup
11/11/22
15:15:45
Detector Setup  (1/2)
11/11/22
15:15:45
Detector Setup  (2/2)
Oven #  1  Det #  1  ( TCD ) Oven #  1  Det #  1  ( TCD )
> - Filtering const  Off > - Auto gain setting  None
 - Sample rate  20 ms  - Full scale value  0
 - Square root calc req  * * * * *  - Auto zero  Off
 - Flame detect level  * * * * *
 - Auto zero value  0.0000  mV
 - Sense set  Equal  - Auto zero time  0.0  s
 - Input resistor  * * * * * * * * * *
Menu Oven# Det# Menu Oven# Det#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.62 
Example of the detector setup screen
F1 (Menu):  Displays the Table Menu screen.
F5 (Oven #):  Sets an oven number.
F6 (Det #):  Sets a detector number.
 TIP
•  The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
•  If the Detector type has been set to None on the Detector Set screen, all the options are shown as * 
(asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type None
TCD
FID
FID-MC
FPD
Filtering const  0.001 1.000
Sample rate  * 20 ms
40 ms
 80 ms
160 ms
Square root 
calc req
 The value is shown except 
when the FPD is installed.
No
Yes
* The square root of the area 
value
Flame detect 
level
 –20 20 mV The item is displayed only when 
the FID, FPD or FID-MC is 
installed.
Sense set  No amplifi cation
Tenfold
Input resistor  1 Mohm
6 Mohm
20 Mohm
60 Mohm
200 Mohm
600 Mohm
1 Gohm
2 Gohm
*:  The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00