Instruction Manual

<5. EtherLCD>
5-43
IM 11B08A01-01E
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Speci es an oven number.
F6 (Det #): Speci es a detector number.
TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
The detector numbers set to Not provided on the Detector Set screen cannot be selected.
The initial screen displays the smallest oven number among those assigned to isothermal
ovens on the Temperature Control Set.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Carrier gas
number
Carrier gas number ---
1-1
1-2
2-1
2-2
3-1
3-2
* --- indicates not used.
Vent-D ow
rate
Alpanumeric: 10 characters
* The detector type is
displayed.
TCD/FID/FID-MC/FPD
Detector type None
TCD
FID
FID-MC
FPD
Vent-REF Alpanumeric: 10 characters
11/11/22
15:15:45
Operating Parameters (4/8)
Oven # 1 Vent # 1
> - Carrier gas number - - -
- Vent type - - -
- Vent owrate
Menu Oven# Vent#
F1 F2 F3 F4 F5 F6
Figure 5.56
Example of the operating parameters screen (4/8)
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Speci es an oven number.
F6 (Vent #): Speci es a vent number.
TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
Only carrier gas numbers for which a gas type is set on the Operation Parameters screen (1/8) can be
selected.
2nd Edition : May 11, 2012-00
5