Instruction Manual
<5. EtherLCD>
5-41
IM 11B08A01-01E
5.4.2 Operating Parameters
11/11/22
15:15:45
Operating Parameters (1/8)
Oven # 1 Carrier gas # 1
> - Carrier gas type - - -
- Carrier gas pressure 0.0 kPa
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.53
Example of the operating parameters screen (1/8)
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Specifi es an oven number.
F6 (Gas #): Specifi es a carrier gas number.
TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• The initial screen displays the smallest oven number among those assigned to isothermal
ovens on the Temperature Control Set.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Carrier gas # Carrier gas number 1 2
Carrier gas
types
Carrier gas type * ---
H
2
He
Ar
N
2
* --- indicates not used.
Carrier gas
pressure
Carrier gas pressure 0
0.0
490
72.5
kPa
psi
Numeric entry
psi (see the defi nition of SI units)
500 kPa = 72.5 psi
*: The settings are valid after rebooting or turning off/on its power.
11/11/22
15:15:45
Operating Parameters (2/8)
Oven # 1 Utility gas # 1
> - Detector number - - -
- Utility gas function - - -
- Utility gas type - - -
- Utility gas pressure 0.0 kPa
- Utility gas fl owrate
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.54
Example of the operating parameters screen (2/8)
2nd Edition : May 11, 2012-00
5