5.4.2 Operating Parameters
11/11/22
15:15:45
Operating Parameters  (1/8)
Oven #  1  Carrier gas #  1
> - Carrier gas type  - - -
 - Carrier gas pressure  0.0  kPa
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.53 
Example of the operating parameters screen (1/8)
F1 (Menu):  Displays the Table Menu screen.
F5 (Oven #):  Specifi es an oven number.
F6 (Gas #):  Specifi es a carrier gas number.
 TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
•  The initial screen displays the smallest oven number among those assigned to isothermal 
ovens on the Temperature Control Set.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Carrier gas # Carrier gas number 1 2
Carrier gas 
types
Carrier gas type  * ---
H
2
He
Ar
N
2
* --- indicates not used.
Carrier gas 
pressure
Carrier gas pressure  0
0.0
490
72.5
kPa
psi
Numeric entry
psi (see the defi nition of SI units)
500 kPa = 72.5 psi
*:  The settings are valid after rebooting or turning off/on its power.
11/11/22
15:15:45
Operating Parameters  (2/8)
Oven #  1  Utility gas #  1
> - Detector number  - - -
 - Utility gas function  - - -
 - Utility gas type  - - -
  - Utility gas pressure  0.0  kPa
 - Utility gas fl owrate  
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.54 
Example of the operating parameters screen (2/8)
2nd Edition : May 11, 2012-00
5