Instruction Manual

<5. EtherLCD>
5-41
IM 11B08A01-01E
5.4.2 Operating Parameters
11/11/22
15:15:45
Operating Parameters (1/8)
Oven # 1 Carrier gas # 1
> - Carrier gas type - - -
- Carrier gas pressure 0.0 kPa
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.53
Example of the operating parameters screen (1/8)
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Speci es an oven number.
F6 (Gas #): Speci es a carrier gas number.
TIP
The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
The initial screen displays the smallest oven number among those assigned to isothermal
ovens on the Temperature Control Set.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Carrier gas # Carrier gas number 1 2
Carrier gas
types
Carrier gas type * ---
H
2
He
Ar
N
2
* --- indicates not used.
Carrier gas
pressure
Carrier gas pressure 0
0.0
490
72.5
kPa
psi
Numeric entry
psi (see the de nition of SI units)
500 kPa = 72.5 psi
*: The settings are valid after rebooting or turning off/on its power.
11/11/22
15:15:45
Operating Parameters (2/8)
Oven # 1 Utility gas # 1
> - Detector number - - -
- Utility gas function - - -
- Utility gas type - - -
- Utility gas pressure 0.0 kPa
- Utility gas owrate
Menu Oven# Gas#
F1 F2 F3 F4 F5 F6
Figure 5.54
Example of the operating parameters screen (2/8)
2nd Edition : May 11, 2012-00
5