Instruction Manual
<5. EtherLCD>
5-40
IM 11B08A01-01E
5.4.1 System Settings
11/11/22
15:15:45
System Settings (1/2)
11/11/22
15:15:45
System Settings (2/2)
> - Analyzer name abc > - Auto start setup Unexecuted
- Analyzer ID 001 - #1 Auto start delay 0 s
- Tag number tag-123
- #2 Auto start delay 0 s
- KGC number KGC00001
- #3 Auto start delay 0 s
- Current date 2011/11/22
- Current time 15:15:45
- Language English
Menu Menu
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.52
Example of the system settings screen
F1 (Menu): Displays the Table Menu screen.
TIP
• The date and time displayed at the top left of the screen are updated when the Current date and Current time
are set.
• The KGC number is shown as a fi ve-digit number with zero padding. For example, if the value is 8924, the
KGC number is displayed as KGC08924.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Analyzer
name
Alphanumeric: 8 characters
Analyzer ID 1 240
Tag number Tag number Alphanumeric: 16 characters
KGC Number 1 99999 The fi xed string KGC is attached
as a prefi x.
Current date Analysis start year
Analysis start month
Analysis start day
2000
1
1
2099
12
31*
Year
Month
Day
In four digits
* Subject to change depending
on the month.
Current time Analysis start hour
Analysis start minute
0
0
23
59
Hour
Minute
Language English, Japanese
Auto start
setup
* Off, On
Auto start
time
1st auto-start time
2nd auto-start time
3rd auto-start time
* 0 10000 Seconds
*: The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00