Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type FID-MC
Det operation 0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
Reset status 0: Initialized
1: Setting done
If the voltage value is zero,
Initialized is displayed;
otherwise Setting done.
Detector type: FID-MC – Execution screen
11/11/22
15:15:45
Det Adjustment
Oven # 1 Det # 1 ( FID-MC )
> - Det operation Off
MC voltage value 0.0 V
- Target MC current
Main << < > >>
F1 F2 F3 F4 F5 F6
Figure 5.47
Example of the detector type: FID-MC – execution screen
F2 (Main): Displays the Det Adjustment (Reset) screen.
F3 (<<): Decreases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).
F4 (<): Decreases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).
F5 (>): Increases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).
F6 (>>): Increases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).
• To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent
key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the
bottom.
• If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255
and the command is successfully sent internally, “Setting succeeded” is displayed. Users
can confi rm the results with the voltage value and current value refreshed every second.
• Selecting the Target MC voltage displays Target MC voltage (AA.A/B.B):. If the value
17.5/0.5 is entered, it will be shown as 17.5+/–0.5 V, 5.0 as 5.0+/– V and 2/1 as 2.0+/– 1.0 V.
• The numbers of ovens to which no detectors are installed cannot be set via F5.