Instruction Manual
<5. EtherLCD>
5-35
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type TCD
Det operation Off, On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
Voltage value 0.0 20.0 V
Current value 0.0 500.0 mA
Target current Numeric, period “.” and slash “/”
are enabled: 16 characters
Input format: AA.A/B.B
Output format: AA.A+/–B.B mA
Detector type: FID-MC - Reset screen
Press F2 (Adjust).
11/11/22
15:15:45
Det Adjustment
Oven # 1 Det # 1 ( FID-MC )
> - Det operation Off
- Reset status Initialized
Menu Adjust Reset Oven# Det#
F1 F2 F3 F4 F5 F6
Figure 5.46
Example of the detector type: FID-MC - reset screen
F1 (Menu): Displays the Table Menu screen.
F2 (Adjust): Displays the Det Adjustment (Execution) screen.
F4 (Reset): Initializes a voltage value.
Pressing F4 displays the options “Abort” and “Initialize.” If the Set/Ent key
is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.
F5 (Oven #): Sets an oven number.
F6 (Det #): Sets a detector number.
• Pressing F4 displays the options “Abort” and “Initialize” in the second line on the bottom. If
the Set/Ent key is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.
• The MC voltage value is read at a one-second interval, when, if the value is 0x0, the status
is refreshed with “Initialized”; otherwise Setting done.
• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.
2nd Edition : May 11, 2012-00