Instruction Manual

<5. EtherLCD>
5-34
IM 11B08A01-01E
Detector type: TCD - Execution screen
Press F2 (Adjust).
11/11/22
15:15:45
Det Adjustment
Oven # 1 Det # 1 ( TCD )
> - Det operation Off
Voltage value 0.0 V
Cuttent value 0.0 mA
- Target current
Main << < > >>
F1 F2 F3 F4 F5 F6
Figure 5.45
Example of the detector type: TCD - execution screen
F2 (Main): Displays the Det Adjustment (Reset) screen.
F3 (<<): Decreases the value by 16 steps within the range of 0 to 255 (for the TCD, by
approximately 0.94 V within the range of 5.0 V to 20.0 V).
F4 (<): Decreases the value by 1 step within the range of 0 to 255 (for the TCD, by
approximately 0.06 V within the range of 5.0 V to 20.0 V).
F5 (>): Increases the value by 1 step within the range of 0 to 255 (for the TCD, by
approximately 0.06 V within the range of 5.0 V to 20.0 V).
F6 (>>): Increases the value by 16 steps within the range of 0 to 255 (for the TCD, by
approximately 0.94 V within the range of 5.0 V to 20.0 V).
To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent
key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the
bottom.
If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255
and the command is successfully sent internally, “Setting succeeded” is displayed. Users
can con rm the results with the voltage value and current value refreshed every second.
Selecting the Target current displays Target current (AAA/BB):. If the value 250/10 is
entered, it will be shown as 250+/–10 mA, 250 as 250+/– mA and 50/5 as +/–5 mA.
The numbers of ovens to which no detectors are installed cannot be set via F5.
TIP
The Det operation, Voltage value and Current value are refreshed at a one-second interval.
2nd Edition : May 11, 2012-00