  Detector type: TCD - Execution screen
Press F2 (Adjust).
11/11/22
15:15:45
Det Adjustment 
Oven #  1  Det #  1  ( TCD )
> - Det operation  Off
Voltage value  0.0 V
Cuttent value  0.0 mA
 - Target current  
Main << < > >>
F1 F2 F3 F4 F5 F6
Figure 5.45 
Example of the detector type: TCD - execution screen
F2 (Main):  Displays the Det Adjustment (Reset) screen.
F3 (<<):  Decreases the value by 16 steps within the range of 0 to 255 (for the TCD, by 
approximately 0.94 V within the range of 5.0 V to 20.0 V).
F4 (<):  Decreases the value by 1 step within the range of 0 to 255 (for the TCD, by 
approximately 0.06 V within the range of 5.0 V to 20.0 V).
F5 (>):  Increases the value by 1 step within the range of 0 to 255 (for the TCD, by 
approximately 0.06 V within the range of 5.0 V to 20.0 V).
F6 (>>):  Increases the value by 16 steps within the range of 0 to 255 (for the TCD, by 
approximately 0.94 V within the range of 5.0 V to 20.0 V).
•  To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent 
key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the 
bottom.
•  If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255 
and the command is successfully sent internally, “Setting succeeded” is displayed. Users 
can confi rm the results with the voltage value and current value refreshed every second.
•  Selecting the Target current displays Target current (AAA/BB):. If the value 250/10 is 
entered, it will be shown as 250+/–10 mA, 250 as 250+/– mA and 50/5 as +/–5 mA.
•  The numbers of ovens to which no detectors are installed cannot be set via F5.
 TIP
The Det operation, Voltage value and Current value are refreshed at a one-second interval.