Instruction Manual
  <5. EtherLCD>
5-33
IM 11B08A01-01E
5.3.9 Detector Adjustment
  Detector type: TCD - Reset screen
11/11/22
15:15:45
Det Adjustment 
Oven #  1  Det #  1  ( TCD )
> - Det operation  Off
 - Reset status  Initialized
Menu Adjust Reset Oven# Det#
F1 F2 F3 F4 F5 F6
Figure 5.44 
Example of the detector type: TCD - reset screen
F1 (Menu):  Displays the Table Menu screen.
F2 (Adjust):  Displays the Det Adjustment (Execution) screen.
F4 (Reset):  Initializes voltage values.
  Pressing F4 displays the options Abort and Initialize. If the Set/Ent key is 
pressed after the selection of Initialize, the TCD voltage value is set to its 
minimum value 5.0 V.
F5 (Oven #):  Sets an oven number.
F6 (Det #):  Sets a detector number.
•  The TCD voltage value is read at a one-second interval, when, if the value is 0.0 V, the 
status is refreshed with Initialized; otherwise Setting done.
•  The Det operation is refreshed at a one-second interval.
•  The numbers of ovens to which no detectors are installed cannot be set via F5.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type TCD
Det operation  0: Off
1: On
* For this operation, the user 
level C or higher is required and 
the operation mode must be 
Stop in Manual status.
Reset status  0: Initialized
1: Setting done
* If the voltage value is 
zero, Initialized is displayed; 
otherwise Setting done.
2nd Edition : May 11, 2012-00