Instruction Manual
  <5. EtherLCD>
5-32
IM 11B08A01-01E
5.3.8  Auto Tuning Operation
11/11/22
15:15:45
Auto Tuning Operation 
 Auto Tuning  Stopping
> - Oven number  ?
 - SYS number  0
 - SYS Method  0
Menu Start Stop
F1 F2 F3 F4 F5 F6
Figure 5.43 
Example of the auto tuning operation screen
F1 (Menu):  Displays the Operation Menu screen.
F3 (Start):  Starts auto tuning.
F4 (Stop):  Ends auto tuning.
 TIP
•  For this operation, the user level C or higher is required.
•  The operation is possible when all the GCMs are in Manual status and the operation mode is Stop. In 
addition, the temperature controller of the associated oven needs to be On.
•  During auto tuning, the oven number, SYS number and SYS method cannot be changed.
•  After "Stopping" message is outputted, temperature may exceed preset temperature temporarily.
•  At the start of auto tuning, a validation check is implemented on the oven number and the 
SYS number and method. If any invalidity is found, auto tuning will not be executed. The 
screen then displays the message “The setting is different.”
* If the SYS number does not belong to the specifi ed oven, it results in an error.
* Even when both the SYS number and method are zero, auto tuning will still be executed.
•  If the temperature controller of the oven is off when auto tuning is started by pressing F3 
(Start), the screen displays the error message “Cannot start auto tuning.”
During auto tuning, pressing F1 (Menu) will not navigate to the Operation Menu screen. It 
displays the message “Cannot operate because of Auto Tuning.”
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Auto tuning 0: Stopping
1: Executing
* If the operation mode for the 
SYS number is Auto Tuning, 
“Executing” is displayed.
Oven number  13
SYS number  (0), 1 6 SYS numbers that are not 
registered in the GCM cannot 
be entered.
SYS method  (0), 1 6
2nd Edition : May 11, 2012-00