Instruction Manual
  <5. EtherLCD>
5-27
IM 11B08A01-01E
F3 (Edit):  Displays the User Program Edit screen.
F5 (Stop):  Executes the forced stop command.
•  After selecting a SYS number, the option displayed is Enable if the execution status is 
currently disabled, and Disable if it is currently enabled.
•  Pressing F5 (Stop) displays Input script number (SYS #). Entering the target script number 
for the end of peak detect executes the forced stop command.
 TIP
The Status is refreshed automatically at a fi xed interval (1 second).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Enable  Disable
Enable
Status  Stopping
Executing
 User Program Edit screen
Press F3 (Edit).
11/11/22
15:15:45
User Program Edit 
Type #1  NO #1  
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
 - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
> -  AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
Main Save Add Del Type# NO#
F1 F2 F3 F4 F5 F6
Figure 5.32 
Example of the user program edit screen
F1 (Menu):  Displays the Operation Menu screen.
F2 (Save):  Saves the script fi le.
F3 (Add):  Add a command line.
F4 (Del):  Delete a command line.
F5 (Type #):  Specifi es a type number.
F6 (No #):  Specifi es a SYS number for the end of peak detect script type; or a period script 
number for the period script type.
 TIP
•  For this operation, the user level C+ is required.
•  The following values can be entered:
  0 to 9, A to Z, ( ) * / + – =, space and backspace.
• Type number
  1: End of peak detect script
  2: Period script
2nd Edition : May 11, 2012-00
5