Instruction Manual

<5. EtherLCD>
5-21
IM 11B08A01-01E
TIP
Display data are not refreshed automatically.
For the operation, the user level C or higher is required.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
StrV # Stream valve
number
131
Stream valve
0 99999999 Times
Operation Status screen (Detector ON Time)
Press F5 (Det).
11/11/22
15:15:45
Operation Status
Oven #1 Det #1
> - 1st action time 0 h
- 2nd action time * * * * * * * *
Menu GC Oven StrV Oven# Det#
F1 F2 F3 F4 F5 F6
Figure 5.24
Example of the operation status screen (Detector ON Time)
F1 (Menu): Displays the Table Menu screen.
F2 (GC): Displays the Total Operation Time screen.
F3 (Oven): Displays the Valve Operation Count screen.
F4 (StrV): Displays the Stream Valve Operation Count screen.
F5 (Oven #): Speci es an oven number.
F6 (Det #): Speci es a detector number.
TIP
Display data are not refreshed automatically.
If the detector type is TCD or FID, the 2nd action time will be shown as * (asterisk).
If no detectors are installed, the data will be shown as * (asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
1st action
time, 2nd
action time
0 99999999 Hours
2nd Edition : May 11, 2012-00