Instruction Manual

<5. EtherLCD>
5-17
IM 11B08A01-01E
Alarm Historical Record
To display the alarm history, press F2 (History).
11/11/22
15:15:45
Alarm Historical Record
Total 1 GCM # 0
Date Time ALM # Contents
11/11 14:10 31 Card ID error 10
Menu Auto Clean GCM#
F1 F2 F3 F4 F5 F6
Figure 5.19
Example of the alarm historical record
F1 (Alarm): Displays the Alarm Status screen.
F3 (Auto): Starts to refresh the alarm status automatically at an interval of 1 second.
F5 (Clean): Flushes the alarm history buffer and starts the Auto refresh. For this operation,
the user level C or higher is required.
F6 (GCM #): Speci es a GCM number.
TIP
The newer data are shown in upper rows in order.
Auto refresh is executed as long as the top data is the newest upon the occurrence of an alarm incident.
Between the Date Time and ALM #, the alarm status is shown as 1 (On) or 0 (Off).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Total Total number of
alarms in the history
data table
GCM # GCM number 0 6 0: Alarm for the entire unit
Date Time Date Time: Month
Date Time: Day
Date Time: Hour
Date Time: Minute
1
1
00
00
12
31
23
59
Month
Day
Hour
Minute
ALM # Alarm number 1 600 Refer to “7.1.4 Alarm
Contents”
Contents
Alphanumeric: 12 characters
* The rst 12 characters are
shown when the length
exceeds the limit.
Alarm information 1 0 0xFFFF
Alarm information 2 0 0xFFFF
2nd Edition : May 11, 2012-00
5