Instruction Manual

<5. EtherLCD>
5-9
IM 11B08A01-01E
F6 (Slot #): Sets a slot number.
TIP
The Status is refreshed automatically at a xed interval (1 second).
On the initial screen, the smallest slot number among those of the D/O cards inserted is displayed.
For the slot number setting, only slot numbers of D/O cards can be accepted.
Up to CH3 are displayed for DI/O cards.
One GCM stream identifying signal can be applied to multiple D/O cards.
In the case where the Stream identi cation has been set on the GCM Setup screen, the D/O is used for that
purpose.
Example: Stream identi cation setup for the GCM1 (where the rst number of the stream
identi cation D/O is 1)
Slot 1 2
Stream
identi cation
DO1 DO2 DO3 DO4 DO5 DO1
Up to 3 streams
Stream
identifying ag
Stream
identi cation
Stream
identi cation
Normal output Normal output Normal output
Up to 7 streams
Stream
identifying ag
Stream
identi cation
Stream
identi cation
Stream
identi cation
Normal output Normal output
Up to 15 streams
Stream
identifying ag
Stream
identi cation
Stream
identi cation
Stream
identi cation
Stream
identi cation
Normal output
Up to 31 streams
Stream
identifying ag
Stream
identi cation
Stream
identi cation
Stream
identi cation
Stream
identi cation
Stream
identi cation
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # Slot number 1 5 Absolute slot number
DO # D/O contact number 1 5 Number Relative contact number
Output item No assignment
Stream sequence
Stream
Operation mode
Alarm
Timing
Cal/Val
Str valve select
* See Note.
Status Off, On
Note:
1: The selected GCM number and stream sequence number are displayed.
2: The selected stream number is displayed
3: The selected GCM number and operation mode are displayed.
4: The selected GCM number and alarm level/concentration alarm are displayed.
5: The selected SYS number, SYS method and sequence order selection are displayed.
6: The selected GCM number and calibration/validation number are displayed.