Instruction Manual
  <5. EtherLCD>
5-7
IM 11B08A01-01E
5.2.3 D/I Status
The D/I (contact input) status is displayed on the screen.
11/11/22
15:15:45
D/I Status 
Slot # 1
DI#  Name  Value  Processing name
 1  a1  Open  Stream sequence
 2  Open  No assignment
 3  Open  No assignment
 4  Open  No assignment
 5  Open  No assignment
 6  Open  No assignment
 7  Open  No assignment
 8  Open  No assignment
Menu Table Slot#
F1 F2 F3 F4 F5 F6
Figure 5.7 
Example of the D/I status screen
F1 (Menu):  Displays the Status Menu screen.
F4 (Table):  Displays the D/I Setup screen.
F6 (Slot #):  Sets a slot number.
 TIP
•  The Value data are refreshed automatically at a fi xed interval (1 second).
•  On the initial screen, the smallest slot number among those of the D/I cards inserted is displayed.
•  For the slot number setting, only slot numbers of D/I or DI/O cards can be accepted.
•  Up to CH3 are displayed for DI/O cards.
If no D/I cards are inserted, the Status Menu screen displays “Not load” on the bottom.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # Slot number 1 5 Absolute slot number
DI # D/I contact number 1 8 Relative channel number
Name D/I name 8 single-byte alphanumeric 
characters
Status Status Open/Close
Processing 
name
Processing name No assignment, Stream 
sequence, Stream, Range 
change Cal/Val, Operation 
mode change, Alarm process
2nd Edition : May 11, 2012-00
5