Instruction Manual
  <5. EtherLCD>
5-6
IM 11B08A01-01E
5.2.2 A/O Status
To display the Analog Output (A/O) status, move the cursor (>) with   or   to the A/O Status 
row and press 
.
11/11/22
15:15:45
A/O Status 
Slot # 1
AO#  Output item  Value
 1  Det 1-1 AO Chromatogram  0.000
  2     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  3     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  4     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  5     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  6     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  7     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
  8     - - - - - -   - - - - - - - -  - - - - - - - - - - - -      0.000
Menu Table Slot#
F1 F2 F3 F4 F5 F6
Figure 5.6 
Example of the A/O status screen
F1 (Menu):  Displays the Status Menu screen.
F4 (Table):  Displays the A/O Setup screen.
F6 (Slot #):  Sets a slot number (applicable only when multiple analog output cards are 
inserted).
 TIP
•  The Value data are refreshed automatically at a fi xed interval (1 second).
•  On the initial screen, the smallest slot number among those of the A/O cards inserted is displayed.
•  For the slot number setting, only slot numbers of A/O cards can be accepted.
If no A/O cards are inserted, the Status Menu screen displays “Not load” on the bottom.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # Slot number 1 5 Absolute slot number
AO # A/O channel number 1 8 Relative channel number
Stream number on 
the output
131
Peak number on the 
output
1 999 Relative peak number
Output item A/O output peak 
name
8 single-byte alphanumeric 
characters
Analysis results on 
the output
0.000
0.000
999.999
9999.999
When the unit is %
For other units
* The number of effective digits 
is 6.
Analysis value unit 
on the output
See the Measuring unit of the 
Peak Setup-Specifi c screen.
* Only the upper 5 characters 
are displayed.
Detector number None, 1-1, 1-2, 2-1, 2-2, 3-1, 3-2
* Oven number and relative 
detector number
Value A/O value -0.25 1.25 If the value stored in the 
database is 10000, it is shown 
as 1.0000.
2nd Edition : May 11, 2012-00