Instruction Manual
  <5. EtherLCD>
5-5
IM 11B08A01-01E
5.2.1 A/I Status
To display the Analog Input (A/I) status, move the cursor (>) with   or   to the A/I Status row 
and press 
.
11/11/22
15:15:45
A/I Status 
Slot # 1
AI#  Name  Value
 1  a1  0.000
 2  0.000
 3  0.000
 4  0.000
Menu Table Slot#
F1 F2 F3 F4 F5 F6
Figure 5.5 
Example of the A/I status screen
F1 (Menu):  Displays the Status Menu screen.
F4 (Table):  Displays the A/I Setup screen (see the 5.4.18 A/I Setting).
F6 (Slot #):  Sets a slot number (applicable only when multiple analog input cards are 
inserted).
 TIP
•  On the initial screen, the smallest slot number among those of the A/I cards inserted is displayed.
•  For the slot number setting, only slot numbers of A/I cards can be accepted.
•  The Value data are refreshed automatically at a fi xed interval (1 second).
If no A/I cards are inserted, the Status Menu screen displays “Not load” on the bottom.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # Slot number 1 5 Absolute slot number
AI # A/I channel number 1 4 Relative channel number
Name A/I name 8 single-byte alphanumeric 
characters
Value A/I value -0.25 1.25 If the value stored in the 
database is 10000, it is shown 
as 1.0000.
2nd Edition : May 11, 2012-00