Instruction Manual
    <4. GC-HMI (touch panel)>
4-41
IM 11B08A01-01E
  Signal:    Detector signal in mV.
Pressing the 
 displays the following information:
  Input signal:  The analog signal from the detector, which has been sampled every 40 
msec, converted to digital values, and then averaged by the given sample 
rate
  Filtered signal:  Input signal, which has been fi ltered using the fi lter constant set as a 
detector signal parameter
  Standard deviation of signal:  Standard deviation of input signal for the past 20 inputs
  Applied voltage:  Bridge voltage of TCD (for TCD only)
  Current:    Bridge current of TCD (for TCD only)
  Flame detection level:  Set value of fl ame detection level (FID, FID-MC, FPD)
  Thermocouple signal:  Thermocouple signal (FID, FID-MC, FPD)
  Flame detection status:  Flame detection status (now burning/burning stopped) (FID, FID-
MC, FPD)
  Methanizer voltage:  Voltage of methanizer (for FID-MC only)
Temperature information
  Indicates the oven temperature information (OVEN, LSV, FPD) of the selected SYS.
  Unit: deg C and deg F.
Temperature controller information
  Indicates the SV value, PV value, and ON/OFF status for the isothermal oven, LSV, and 
FPD, of the selected SYS.
 Press 
.
Figure 4.54
Pressure information
  Displays pressure information of EPC (Carrier n-1 to n-2, and Utility n-1 to n-4 (where n is 
oven number)) inthe selected oven.
2nd Edition : May 11, 2012-00
4