Instruction Manual
    <4. GC-HMI (touch panel)>
4-30
IM 11B08A01-01E
Cal (Val):  Executes calibration or validation of the specifi ed number. After the 
selection, set the calibration (validation) number (1 to 3 for calibration, or 
1 to 3 for validation). Ensure that the calibration (validation) number is the 
stream number.
Start cal (val):  Starts the calibration (validation).
Finish cal (val):  Finishes the calibration or validation.
Cancel spec.:  Cancels the parameter settings for changing the measurement status.
Cal (val) method:  Set the desired mode of calibration or validation: manual, semi-auto, or 
auto.
Figure 4.36
Auto:  Calibration or validation is executed automatically according to the set 
values of start time and time interval.
Semi-auto:  When a calibration or validation number is specifi ed, the stream valve 
on the calibration or validation stream automatically opens, followed by 
calibration or validation.
Manual:  When a calibration or validation number is specifi ed, all stream valves 
close, allowing a standard sample to fl ow. Then, calibration or validation 
can be performed manually.
Start/stop auto cal:  Starts/stops automatic calibration.
2nd Edition : May 11, 2012-00