Instruction Manual
    <3. Basic Operation and Startup>
3-29
IM 11B08A01-01E
11/03/22
15:15:45
Peak Setup-General
GCM #1 
> -  Conc calc req    Standard
  -  Det slope ratio    1.00
 - Calib range    0.500
  -  Calib repeat range  2.000
  -  Calib coef range  4.000
 - #1 Tickmark setup  None
 - #2 Tickmark setup  None
  -  Peak detect err set  Executed
Menu Gate Valve Atm DO GCM#
F1 F2 F3 F4 F5 F6
Figure 3.24 
Example of Common peak setting screen
 The method of changing the Reference area
Setting the Calibration factor range to 0 changes the Calibration factor in the Individual peak 
setting (Figure 3.23) to 1 after the calibration and also changes the Reference area to the 
value obtained by calculation. In this method, values are changed regardless of the value of 
the Reference area obtained, so the automatic calibration fi nishes normally and the calibration 
value is updated even if the measurement result of the standard sample is abnormal. Therefore, 
from the set values of the Calibration range and the Calibration repetition range, decide by the 
following method if the calibration result should be refl ected.
When the Calibration range has been set to  and the measurement result of the standard 
sample is out of the Reference concentration ± the Measuring range × , an alarm message 
appears and the calibration result is not refl ected.
When the Calibration repetition range has been set to  and the variation coeffi cient obtained 
from n times of measurement results of the standard sample is larger than , an alarm message 
appears and the calibration result is not refl ected. A variation coeffi cient is a value obtained by 
the division of a standard deviation by a measuring range. For the equation, see the Appendix E 
Calculation.
This method is not suitable for the history management of calibration. A method of changing the 
Calibration factor, described below, is recommended to normal calibration. 
At the factory, calibration has been conducted according to the method of changing the 
Reference area. 
 The method of changing the Calibration factor
Setting the Calibration factor range to other than 0 does not change the Reference area in the 
Individual peak setting after the calibration (Figure 3.23) but changes the Calibration factor to the 
value obtained by calculation. 
In this method, decide by the following method if the calibration result should be refl ected from 
the Calibration factor. At the same time, use the Calibration range and the Calibration repetition 
range.
When the Calibration factor range is set to  and the obtained Calibration factor is between 1 
–  and 1 + , the calibration result is refl ected. If the obtained Calibration factor is smaller than 
1 –  or larger than 1 + , an alarm message appears and the calibration result is not refl ected. 
For example, when the Calibration factor range is set to 0.1 and the obtained Calibration factor is 
between 0.9 and 1.1, the calibration result is refl ected.
2nd Edition : May 11, 2012-00
3










