Instruction Manual

xiv
IM 11B08A01-01E
Installation site and environment
The analyzer speci cations allow it to be used in hazardous areas as de ned by DIV1, GPS B, C
& D, T1, T2, T3, T4 (FM) or Zone 1 IIB + H2T1, T2, T3, T4 (ATEX, IECEx, TIIS). However, never
install the analyzer in an area where dense explosive gas exists for a long time.
For the class of hazardous areas:
For FM, refer to Article 500 of the National Electrical Code (NEC).
For ATEX or IECEx, see IEC 60079-10-1.
For TIIS explosion protection, refer to Article 1 (15) to (17) of the Constructional Requirements for
Electrical Equipment for Explosive Atmospheres (Japanese only).
Wiring works
<FM>
WARNING
All wiring shall comply with National Electric Code ANSI/NFP A 70 and Local Electric Codes.
In a hazardous area, use conduits for wiring in the protection system or to electronics
sections.
CAUTION
The unused electrical connection ports should be closed with an appropriate ameproof-
certi ed plug.
Analyzers have pressurized enclosures. The cable end should be sealed in order to
maintain pressure to the pressurized enclosure. Otherwise, power is not supplied to the
electronics section.
<ATEX, IECEx>
WARNING
All wiring shall comply with Local Electric Codes and Requirements.
In a hazardous area, use appropriate ameproof-certi ed parts for connecting cables.
All externally powered input signals into the pressurized enclosure shall be isolated by
external relays controlled by the Ex px protection system(Ex px safety device).
CAUTION
The unused electrical connection ports should be closed with an appropriate ameproof-
certi ed plug.
Analyzers have pressurized enclosures. The cable end should be sealed in order to
maintain pressure to the pressurized enclosure. Otherwise, power is not supplied to the
electronics section.
2nd Edition : May 11, 2012-00