Instruction Manual
    <2. Installation, Piping, and Wiring>
2-22
IM 11B08A01-01E
2.3.4  Connecting Power Cable and Grounding
 CAUTION
•  Wire the power supply cable keeping the distance of 1 cm or more from other signal wires.
•  The power supply cable shall comply with UL or CSA.
•  Do wiring after securing protective grounding.
Use crimp-on terminals for all power cables and grounding (see Figure 2.18).
Use crimp-on terminals suitable for the cable core (see Table 2.8).
Lug length c
Insulation covering 
inside diameter d
Hole 
diameter
a
Lug outside
diameter b
Figure 2.18  Crimp-on terminal
Table 2.8  Size of crimp-on terminal
Nominal cross 
sectional area
Screw 
(mm)
Hole 
diameter a 
(mm)
Outside 
diameter b 
(mm)
Length c 
(mm)
Insulation 
covering inside 
diameter d (mm)
Applicable 
terminal*
5.5 mm
2
4 4 to 5 9.8 or less 25 to 29 5.8 or less
AMP 170785-1
JST 5.5-4
2.0 mm
2
4 4.3 or more 8.7 or less approx. 21 5.8 or less
AMP 170782-1
JST V2-4
1.25 mm
2
4 4.3 or more 8.7 or less approx. 21 5.8 or less
AMP 170782-1
JST V1.25-4
*:  AMP: Japan AMP Co., Ltd.
  JST: JST Co., Ltd.
 Power supply line to the protection system (A) (B)
The power supply to protection system A is used for both heater power and electric circuit power.
The power supply or protection system B is used only for heater power.
Connect the attached ferrite core for ATEX or IECEx.
Grounding must be wired.
2nd Edition : May 11, 2012-00