Instruction Manual

<2. Installation, Piping, and Wiring>
2-16
IM 11B08A01-01E
3/4NPT(F)
9/16 inch
A
3/4NPT
Connection port of the protection system *3
Connection port of the electronics section *1
Electronic section
(Pressurized enclosure)
*1: Any of the six connection ports of the
electronics section can be equally used.
*2: A plug is attached to connector.
Remove and keep it for future use.
*3: In the FM-Y, the protection system is
not provided.
Protection system *3
(Flameproof enclosure)
as close as possible
not more than 18 inches (457 mm)
18 inches (457mm)
maximum
Sealing Fitting
Conduit
HAZARDOUS
LOCATIONS
NON-HAZARDOUS
LOCATIONS
Detail of A
Connector
Nut
Gasket
Plate
Plug *2
NONHAZARDOUS
LOCATION
EQUIPMENT
NONHAZARDOUS
LOCATION
EQUIPMENT
Figure 2.17 Wiring in FM speci cation
ATEX, IECEx
WARNING
All wiring shall comply with Local Electric Codes and Requirements.
In a hazardous area, use appropriate ameproof-certi ed parts for connecting cables.
All externally powered input signals into the pressurized enclosure shall be isolated by
external relays controlled by the Ex px protection system(Ex px safety device).
2nd Edition : May 11, 2012-00