Instruction Manual
  <1. Overview>
1-86
IM 11B08A01-01E
The following table shows processing designation of peak individual settings for the tracking 
(correction) target.
Processing designation Tracking 
specifi cation
Tracking 
(correction) 
items
Stream type Concentration 
calculation
Type Details
Peak processing External Tertiary No tracking
Tracking 
Standard*1
Gate ON time
Gate OFF time
Calibration, 
Measurement, 
Validation
Standard
External straight 
line
Calibration, 
Measurement, 
Validation
Indirect method Calibration, 
Measurement, 
Validation
Standard, 
Correction area 
fraction
Calculation 
processing
Base level No tracking
Tracking
Operation time Calibration, 
Measurement, 
Validation
*1  “No tracking” or “Tracking” is settable for tracking specifi cation when the gate processing method is “Zone gate.” Only the setting 
for zone gate head peak becomes valid for tracking specifi cation.
11/03/22
15:15:45
Peak Setup-Specifi c (4/5)
Stream #1 Peak #1   External 3rd
> -  Auto tracking req  Tracking std
  -  Peak std time    100.0 s
  -  Tracking coef a    1.0
  -  Tracking coef b    0.0
Menu Copy1 Copy2 Atm Strm# Peak#
F1 F2 F3 F4 F5 F6
Figure 1.44 
Tracking specifi cation and tracking factor a/b in peak individual setting (GC-HMI 
EtherLCD)
 Tracking factor a/b
Tracking factor a/b is a factor to perform correction of gate ON/OFF time and operation time of 
base level processing by straight-line approximation. When the tracking factor b is 0, proportional 
correction by tracking factor a is performed.
--------------------------------------------------------------------------------------------------------------------------------
Gate ON time after tracking = Gate ON time × Tracking factor a + Tracking factor b
Gate OFF time after tracking = Gate OFF time × Tracking factor a + Tracking factor b
Base level operation time after tracking = Base level operation time × Tracking factor a + Tracking 
factor b
--------------------------------------------------------------------------------------------------------------------------------
Tracking factor a/b can be obtained by auto-calculation or by entering a set value.
To obtain tracking factor a/b by auto-calculation, set “Update” for “Automatic renewal of tracking 
factors” of SYS method. Automatic renewal of tracking factor a/b is performed after completion of 
peak detection of all the peaks for the relevant SYS method but before correction of gate ON/
OFF time and operation time of base level processing.
Refer to the following items to fi nd the details of automatic renewal of tracking factor a/b.
2nd Edition : May 11, 2012-00