Instruction Manual

<1. Overview>
1-81
IM 11B08A01-01E
EPC program setting
When pressure adjustor for career gas complies with EPC speci cation of program control,
three-staged heating program can be set.
Time
Warm up time
for stream 2
Stream 1
Start Start
Analysis cycle
Stream 2
Analysis cycle
Pressure
3rd control
pressure
2nd control
pressure
1st control
pressure
Initial control
pressure
1st rise
2nd rise
3rd rise
Initial
control
time
1st
control
time
2nd
control
time
3rd
control
time
Figure 1.39 EPC program operations
11/03/22
15:15:45
SYS Setup (3/4)
SYS #1 Oven #1
- Carrier 1-1 Executed
- Carrier 1-2 Executed
> - Utility 1-1 Executed
- Utility 1-2 Executed
- Utility 1-3 ----
- Utility 1-4 ----
Menu Peak Status Get SYS# Oven#
F1 F2 F3 F4 F5 F6
Figure 1.40
Setting example of SYS (GC-HMI EtherLCD)
2nd Edition : May 11, 2012-00
1