User guide

<4. Engineering>
4-3
IM 11B06E01-01E
(3) ToenableJava,selectGeneraltabandclickUseJavaprogram.
(4) Setting the network
SpecifytheEthernetconnectionforASGWifitissingleorduplex.
EthernetofASGWhasdualorsingletypeanddualtypehasduplexandseparatedmode.
After completing the above 3 items, download the information by [File] - [Download].
ReboottheASGWbypressingResetbuttonorWebbrowser.AfterstartingASGWconrm
thattheIPaddressiscorrectbypingcommandofPC.
(5) AllocatingapplicationstotheDOterminal(onlywhen#DOfunctionisused)
Whenthe#DOsettingisusedtooutputalarminformationtotheDOterminal,theDOvari-
able called device label in the ASGW application must be mapped to the I/O terminal by
usingtheresourcecongurator.Immediatelyaftertheinitialinstallation,thedevicelabelis
not related to the I/O terminal.
TheDOoutputdenitionismatchedtothedefaultfactorysettingofthedevicelabelofthe
system card. If none of the default settings are changed, this procedure is not needed.
(6) Downloading settings
DownloadthevesettingsabovetoASGW.
Resourcecongurator:[File]-[Download]
Afternishing,rebootASGW.UsethepingcommandfromthePCtoconrmthattheIPad-
dress has been set correctly.
(7) Setting the serial port
In the case of Modbus RTU, the communication parameters should be adjusted to a con-
nected device.
Referto4.3.2Modicationofserialportsetting.
4th Edition : Nov. 22, 2011-00