5.4.3 GC8/GC1000 Type Output Data Format (Fixed to 45 Characters)
(1)Analysis value data (concentation)
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
DS
,,
%
PPM A: CHL
A: CLL
A: RT
CR: Carriage return,
LF: Line feed
,,,
T
C
R
L
F
bacd e fgh
a: Stream number (01 to 24) b: Component number (01 to 24) c: Concentration (without zero suppression) d: Unit
e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL) f: Retention time (seconds) g: Retention time error alarm
h: Analyzer number 1 to 9
(Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs)
F0511.EPS
(2)Analysis value data (component ratio operation, linear polynormial operation)
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
D
S
1
2,,
Dummy space
C
R
L
F
b eac
a: Stream number (01 to 24) b: Component number (01 to 24) c: Operation value (without zero suppression) d: Unit e:Analyzer number 1 to 9
F0512.EPS
%
PPM
d
(3)Calibration factor
1 234567 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
CC
,,
,
COE
Dummy space
C
R
L
F
b eacd
a:
Standard sample (STD1=1, STD2=2, STD3=3)
b:
Component number (01 to 24)
c:
Calibration factor (without zero suppression)
d:
Sensitivity error (if an error occurs)
e:
Analyzer number 1 to 9
F0513.EPS
(4)Error alarm
• Memory pattern check error • Watchdog timer
• Calibration out of range • Detector 1 (or 2) calibration error
• Isothermal oven temperature error • Power off
• Pressure switch 1 (or 2) OFF • External contact inputs 1 to 8
• Communication error • FID1 (or 2) extinguished
• Calibration repeatability error
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
AS /:
,,
Dummy space
C
R
L
F
bacde f
a: Month, b: Day, c: Hour, d: Minute e: Error alarm types
f: Analyzer No.1 to 9
MEM: Memory pattern check error
WDT: Watchdog timer
AD1: Detector 1 calibration error
AD2: Detector 2 calibration error
TMPH: Isothermal oven temperature error
CAR1: Pressure switch 1 OFF
CAR2: Pressure switch 2 OFF
EXT1: External contact input 1
EXT2: External contact input 2
EXT8: External contact input 8
FLM1: FID1 extinguished
FLM2: FID2 extinguished
RPT: Calibration repeatability error
CAL: Calibration out of range
POF: Power off
NSD: Communication error
(GC1000 - GCCU)
Note: For 3 characters, the 18th character is a dummy space.
F0514.EPS
8 items in total