<6. Analysis Results Windows>
6-12
IM 11B03G03-03E
l Procedure
(1) Open the Concentration Analysis History or Retention Time History window. Click on the
[Automatic Saving] command in the [File] menu. The Automatic Saving dialog box pops up.
F0610.ai
(2) Click on the [On] radio button in the Automatic Saving eld.
(3) Specify the automatic saving cycle in either a measurement cycle or a specic time for daily
saving in the Saving Cycle eld.
(4) Specify the directory where the data is to be saved in the Automatic Saving Directory eld.
The default setting is “\data” directory right under the installation directory.
(5) Click on the [OK] button. The data of the concentration analysis history and retention time
history will be simultaneously and automatically saved in two les as follows.
Two les are saved with the following names again and again each time analysis ends. (The
les are overwritten every time so that the contents are always updated to the latest history
record data.)
Concentration Analysis History: C_NOW. TMP
Retention Time History: R_NOW. TMP
In addition, the data are automatically saved in new les at every automatic-saving cycle.
The les are saved with a le name including the year (last one digit), month (1 to 9, a, b, c),
day, time information and the appropriate extension.
Assuming that the les are automatically saved after the history record data is acquired at
17:30, December 15, 1999, their le names are:
“9c151730.cnc” and “9c151730.rtm”
The data type is identied by the extension: “.cnc” represents the concentration analysis
history data and “.rtm” represents the retention time history data.
Files to be automatically saved are saved under the directory specied in the Automatic
Saving Directory eld in the Automatic Saving dialog box.
From the next transfer after clicking on [OK] in the dialog box, automatic saving is executed.
Re-specifying the automatic saving cycle resets the number of cumulative analyses to zero.
Counting restarts from zero for the re-specied cycle.
(6) To quit automatic saving, open the Concentration Analysis History or Retention Time History
window, and again click on the [Automatic Saving] command in the [File] menu. Then select
[Off] in the Automatic Saving dialog box.
NOTE
Automatic saving requires a certain free space in the hard disk. If the remaining free space
decreases, the automatic saving must be stopped.