Manual
<6. Analysis Results Windows>
6-7
IM 11B03G03-03E
n RetentionTime History Window
The Retention Time History window is used to view the trend of the retention time data which the
analyzer has analyzed to date in the Process mode. This window shows all the peaks assigned
at the startup of the Analysis Results window. The Retention Time elds of peaks not measured
at the given time remain blank. To view the data not shown on the initial screen, use the scroll bar.
l Window Layout
The following gure shows the layout of the Retention Time History window.
F0607.ai
l Contents of Information
The following table summarizes the contents of the data items shown in the Retention Time
History window.
Item Description
# Indicates the relative number (1 to 255) of peaks (the serial number of all peaks of all
streams).
Stream# - Peak#
- Peak name
Indicates the peaks of the stream, in ascending order (in the order of their assignment)
of the peak number, in which the latest analysis has been executed.
The system uses dashes “-” to connect the stream number, peak number and
peak name in sequence, followed by the concentration unit (%, ppm or no unit) in
parentheses.
Date/time Indicates the date (year/month/day) and time (hour/minute) when the analysis for each
cycle started (up to 250 date/time data items). If the number of data items exceed 250,
the oldest data item is discarded to make way for a new data item.
Retention time Indicates the retention time for each peak (in seconds).
4th Edition : Mar. 21, 2012-00