Manual
    <4. Chromatogram Window>
4-9
IM 11B03G03-03E
4.4  Changing the Scale and Scrolling the 
Display
There are three scales on the chromatogram display: X-axis [sec], Y-axis [mV], and temperature 
[°C]. These scales can be changed to the desired ones using the commands in the [Scale] menu 
or Enlarge/Reduce buttons.
n  Disabling and Enabling the Auto-scale Function
In the Chromatogram window scales are automatically set to display an overall view of the 
chromatogram. This function is called “auto-scale”.
 TIP
Changing the scale automatically disables the auto-scale function.
This section explains how to enable/disable the auto-scale function.
l Procedure
(1)  To disable the auto-scale, click on either the [Auto Scale] command in the [Scale] menu or 
the Auto-scale button on the toolbar. This enables the auto-scale, causing the check mark (
) next to the [Auto Scale] command to disappear and the Autoscale button to appear not-
depressed.
(2)  To enable the auto-scale, click again on either the [Auto Scale] command in the [Scale] 
menu or the Auto-scale button on the toolbar. This enables the auto-scale function again, 
causing the check mark (
) next to the [Auto Scale] command to appear and the Auto-scale 
button to appear depressed.
n  Changing the Scale
l Enlarge/Reduce Buttons
Each clicking on the Enlarge/Reduce buttons changes the scale as shown in the following table.
buttons: reduces the scale
buttons: enlarges the scale
Scale Available Scale Increments
X-axis scale (sec) 0.2→0.5→1→2→5→10→20→50→100→200→500→1000→
2000→5000→10000→20000→50000
Y-axis scale (mV) 0.1→0.2→0.5→1→2→5→10→20→50→100→200→500→1000
Temperature (°C) 10→50→100→200
The scales are common to both chromatograms and graphs of temperature data.
n  Scrolling the Display
Scrolling the display to view the chromatogram not shown on the display, typically when the 
entire chromatogram cannot t in the display because of, for example, scale enlargement.
The display can be scrolled using either of the following two methods.
l Method 1
Click on the  ,  ,  ,   or button at either end of the scroll bars. The chromatogram scrolls in 
the direction of the arrow by one increment of the scale with each click.
4th Edition : Mar. 21, 2012-00