Manual

<3. Analyzer Operation Window>
3-21
IM 11B03G03-03E
n Uploading
To upload and save parameter settings from the analyzer to a disk of the Maintenance Terminal,
take the following procedure.
l Procedure
(1) Click on the [Upload to PC] command in the [File] menu. The Save As dialog box pops up
where you can specify the le name for the parameter setting information to be uploaded.
The default directory is “Data” right under the installation directory.
Input
F0336.ai
(2) Type the name of the le where you want to save the parameter settings, with an extension
of “.pr6” in the “File name” eld.
Example: When the le name is “system01”, type
system01.pr6
(3) Click on the [OK] button. A dialog box pops up to show the status of the transmission of the
parameter settings from the analyzer.
F0337.ai
When all parameter settings have been uploaded and saved in the le, the following dialog
box pop up.
F0338.ai
It takes about 40 seconds for data transmission from the analyzer (if the communication
speed is 38400 bps).
(4) Click on the [OK] button. The uploading has been completed.
HELP !
To interrupt the uploading of parameter settings;
Click on the [Interrupt] button in the dialog box showing the transmission status. A conrmation dialog box
pops up.
Click on the [Yes] button.
4th Edition : Mar. 21, 2012-00