IM 11B03G03-03E
n  Alarm Windows
Message Description Corrective Measures
History 
deletion
Alarm history cannot be 
cleared in User level A.
Invalid attempt to reset the alarm 
history at the wrong user level.
Alarm histories can be reset at 
user level B or C only.
Creation of 
user alarm 
description
Factory alarm 
descriptions cannot be 
changed. Select User 
dened alarm.
A number other than the user 
dened alarm numbers has been 
specied when creating a user 
alarm description.
Conrm the valid alarm numbers, 
and then select a correct alarm 
number.
n  Analysis Result Window
Message Description Corrective Measures
Memory Memory AI location Error. 
Not enough memory.
The memory is too low. Check that memory is reserved 
sufciently for work.
File 
operation
File name is incorrect. 
This le is not XXX.
Invalid attempt to open the le 
that does not contain the data 
corresponding to the onscreen 
window.
Open the le appropriate for the 
on-screen window.
Read error. Data is 
incorrect. Error occurred 
reading a le.
Failed to open the le because it 
contained invalid data.
Open the le that contains the 
valid data.
Error occurred while 
saving le. File cannot be 
saved.
An error occurred while saving 
le. Filed to save the le.
Eliminate the cause of the error 
and then retry saving the le.
Data has been changed.
Do you want to save it?
Changes have been made to the 
data of the window.
If the le needs to be saved 
before the window is closed, 
select “Yes.”
Data 
editing
Entered valve is incorrect. 
Only a numerical value 
can be entered here.
Invalid attempt to type in a data 
value other than numerals.
Type in a numeric data value.
Automatic 
saving
Entered value is incorrect. 
Automatic saving cycle 
must be between 1 and 
250.
An invalid number has been 
typed in as the interval for 
automatic saving.
For the interval for automatic 
saving, type a whole number 
from 1 to 250.
Graph 
creation
Graph cannot be drawn.
No data available.
Cannot create a graph because 
the currently open window does 
not contain graph data.
Create a graph when data are 
available.
Graph cannot be drawn.
Data is incorrect.
Cannot create a graph because 
the selected range for the graph 
being created contains invalid 
data. This message also appears 
if the selected range for the 
graph does not contain any data.
Once again check the data for 
the selected range of graph. Also 
check that the range contains 
necessary data.
Selection error.
Necessary data is not 
selected.
Some of the data necessary 
to create the graph remain 
unselected.
Once again check that none 
of the necessary data remain 
unselected.
Selection error. You can 
select 1 to 6 Peak names.
Seven or more peaks have been 
selected at one time during peak 
selection.
A graph can contain no more 
than six peaks at a time. Select 
no more than six peaks.
Selection error. Start time 
is set after End time.
The starting time has been set at 
an later time than the ending time 
during time-axis setting (manual).
Set the starting time at a time 
earlier than the ending time.
Graph cannot be drawn.
Min./max. axis scale of a 
Logarithm graph cannot 
be set or under zero.
The upper or lower limit of 
the scale has been set at a 
value smaller than zero during 
the setting (manual) of the 
logarithmic axis.
A logarithmic axis cannot indicate 
values smaller than zero.
When setting the upper and 
lower limits of a logarithmic axis, 
use values larger than zero.