Instruction Manual
    <5. OPC DA 2.0 Interface>
5-1
IM 11B06F01-01E
5.  OPC DA 2.0 Interface
ThischapterexplainstheOPCDA2.0interfaceofASGWandASIU.
Adding an OPC server of STARDOM to the system allows ASGW/ASIU to have the OPC in-
terface. There is no need to add any special settings to the ASGW/ASIU application. Map-
ping Modbus data of ASIU to ASGW allows ASIU data to be read and written by accessing 
ASGW through the OPC interface.
5.1 SystemConguration
GC8000
ASIU
ASGW
Ethernet
OPC Server PC
GC1000 MarkII
F0501.ai
5.2  Software Required
ASGW do not require additional software to be installed for OPC. The STARDOM software listed 
below is required for a PC on which an OPC server is to be installed.
Model Item Remark
NT781AJ FCN/FCJ OPC Server for Windows
NT783AJ Redundant Network Function License 
for FCN/FCJ OPC Server
Required only for redundant network.
5.3  Installation and Settings
For installation and settings of the OPC server and the duplexed package, refer to their respec-
tive user’s manuals or online help.
5.4  Update rate
Data for the OPC interface is updated every second. Therefore, a delay of about one second 
may occur when the OPC client reads the latest data from analyzers.
TIP
Separating OPC data from Modbus data causes hogging of the CPU. To avoid this, the OPC data update rate is 
set to 1 second while the Modbus communication is updated every 200 ms.
4th Edition : Nov. 22, 2011-00