User Manual
IM 12B6J8-01E-E
9
2.8 Regulatory standards
CE : Decision 768/2008/EC
- ATEX : Directive 94/9/EC, as amended by Regulation (EC) no.
1882/2003
- Certificate no. : DEKRA 11ATEX0064 X
II 1 G Ex ia IIC T3...T6 Ga
- Electrical data : For sensor input circuits (by connector) connected to a certified
intrinsically safe circuit with the following maximum values
: Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 µH; Ci = 30 µF
::or
: Certified intrinsically safe Yokogawa transmitter
Model FLXA21 series.
- Special
- conditions (X) : T6 for Tamb. -40 °C to +60 °C
T5 for Tamb. -40 °C to +75 °C
T4 for Tamb. -40 °C to +110 °C
T3 for Tamb. -40 °C to +125 °C
: Electrostatic charges on the sensor enclosure shall be avoided.
- Pressure : Directive 97/23/EC, as amended by Regulation (EC) no.
1882/2003
- Applying article : 3.3 (Sound Engineering Practice)
: Damaging the screw thread of the sensor might influence the
maximum process pressure.
- EMC : Directive 2004/108/EC
IEC 61326-1: 2006 Class A (control and laboratory use)
IEC 61326-2-3: 2006 (use in industrial locations)
- Low Voltage : Directive 2006/95/EC
: Sensor contains glass parts which if broken can cause cutting
injuries.
- WEEE : Directive 2012/19/EU
- RoHS : Directive 2011/65/EU
IECEx
Applying standards : IEC 60079-0: 2007
IEC 60079-11: 2006
IEC 60079-26: 2006
Certificate no. : IECEx DEK 11.0065X
Ex ia IIC T3...T6 Ga