Instruction Manual Model FU24F pH/ORP SENCOM™ sensor IM 12B6J8-01E-E 1st edition
(BG) Всички улътвания за продукти от серията ATEX Ex ce лредлагат на английски език. Ако се нуждаете oт улътвания за продукти oт серията Ex на родния ви eзик, ce свържете c най-близкия oфис или представителство на фирма Yokogawa. (CZ) Všechny uživatelské příručky pro výrobky, na něž se vztahuje nevýbušné schváleni ATEX Ex, jsou dostupné v angličtině. Požadujete-li pokyny týkající se výrobků s nevýbušným schválením ve vašem lokálním jazyku, kontaktujte prosím vaši nejbližší reprezentačni kancelář Yokogawa.
(LV) Visas ATEX Ex kategorijas izstrâdâjumu Lietoðanas instrukcijas tiek piegâdâtas angïu valodâs. Ja vçlaties saòemt Ex ierîèu dokumentâciju citâ valodâ, Jums ir jâsazinâs ar firmas Jokogava (Yokogawa) tuvâko ofisu vai pârstâvi. (LT) Visos gaminiø ATEX Ex kategorijos Eksploatavimo instrukcijos teikiami anglø kalbomis. Norëdami gauti priestaisø Ex dokumentacijà kitomis kalbomis susisiekite su artimiausiu bendrovës Yokogawa biuru arba atstovu.
CONTENTS 1. PREFACE 5 2. GENERAL SPECIFICATIONS 7 1.1 1.2 1.3 1.4 Introduction Unpacking and Checking Warranty and Service Serial Number definition 2.1 Measuring elements 2.2 Wetted parts 2.3 Functional specifications (at 25°C) 2.4 Dynamic specifications 2.5 Operating range 2.7 Power supply (Supply+ versus Supply Gnd) 2.8 Regulatory standards 2.9 Shipping details 2.10 Environment and operational conditions 2.11 Mechanical specifications 5 5 6 6 7 7 7 7 8 8 9 12 12 12 3.
1. PREFACE 1.1 Introduction This Instruction Manual provides information for the installation and use of the FU24F SENCOM™ sensor. This digital sensor shows how Yokogawa applies the motto “Simply the Best” to sensor technology. Setup of the sensor is very easy because all sensor specific characteristics, such as calibration data, are stored in the sensor.
1.3 Warranty and Service Yokogawa products are guaranteed free from defects in workmanship and materials under normal use and service for a period of (typically) 12 months from the date of shipment from the manufacturer. Individual Sales organizations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted.
2. GENERAL SPECIFICATIONS 2.1 Measuring elements pH glass electrode Ag/AgCl reference Solid Platinum electrode Pt1000 temperature sensor 2.2 Wetted parts Sensor body Measuring sensor LE glass tube Reference junction Earth pin O-ring Bellow system : : : : : : : PPS 40GF (RytonTM with glass filling) G-glass AR-glass Porous PTFE Solid Platinum Viton Viton 2.
2.5 Operating range pH : 0 to 14 ORP : -1500 to 1500 mV rH : 0 to 100 Temperature - Dome shape : -10 °C to 105 °C (14 °F to 221 °F) - Flat surface : +15 °C to 105 °C (59 °F to 221 °F) Pressure : p(bar) 10 NPT 5 FSM 0 -10 Conductivity 0 25 50 Temp. (ºC) 105 : > 10 µS/cm Note: T he pH operating range is 0-14 pH, but using the sensor at temperature- and / or pH-extremes will seriously shorten the lifetime. Note: Sensor is suitable for pure water applications. 2.
2.8 Regulatory standards CE - ATEX - Certificate no. - Electrical data - Special - conditions (X) : Decision 768/2008/EC : Directive 94/9/EC, as amended by Regulation (EC) no. 1882/2003 : DEKRA 11ATEX0064 X II 1 G Ex ia IIC T3...T6 Ga : For sensor input circuits (by connector) connected to a certified intrinsically safe circuit with the following maximum values : Ui = 6.1 V; Ii = 230 mA; Pi = 1.
CSA Certificate no. Master contract no. : 2516979 : 182892 IS, Class I Div. 1, GP A, B, C, D T3…T6 Electrical data : For sensor input circuits (by connector), connected to a certified intrinsically safe circuit, with the following maximum values : Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF or Certified intrinsically safe Yokogawa transmitter Model FLXA21 series. Ambient temperature : T6 for Tamb. -40 °C to +60 °C T5 for Tamb. -40 °C to +75 °C T4 for Tamb.
FM Certificate no. : 3046277 IS, Class I Div. 1, GP A, B, C, D T3…T6 Electrical data : For sensor input circuits (by connector), connected to a FM approved intrinsically safe apparatus meeting the entity parameters of the SENCOM™ sensor: Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF or FM approved intrinsically safe Yokogawa transmitter Model FLXA21 series. Ambient temperature: T6 for Tamb. -40 °C to +60 °C T5 for Tamb. -40 °C to +75 °C T4 for Tamb. -40 °C to +85 °C T3 for Tamb.
Note: When the sensor has been connected to not intrinsically safe equipment which exceeds the restrictions regarding the sensor input circuit (see electrical data), the sensor is not suitable anymore for intrinsically safe use. 2.9 Shipping details Package size (L x W x H) : 300 x 100 x 75 mm (11.8 x 3.9 x 3.0 inch) Package weight : app. 0.33 kg (0.73 lbs) 2.
4. DIMENSIONS Dimensions in mm (inches) 257.5 ±1.7 (10.14” ±0.07) 192 ±1.1 (7.56” ±0.04”) ø20.4 ±0.05 (ø0.8” ±0.002”) ø28 ±0.2 (ø1.1” ±0.01”) 70 ±0.9 (7.56” ±0.04”) 1-11.5 NPT 1-11.5 NPT 243.5 ±1.7 (9.59: ±0.07”) 178 ±1.1 (7.01” ±0.04”) ø20.4 ±0.05 (ø0.8” ±0.002”) ø28 ±0.2 (ø1.1” ±0.01”) 56 ±0.9 (2.2” ±0.04”) 1-11.5 NPT 1-11.
5. WIRING The FU24F is provided with a multipole M9 male connector for connection to the Yokogawa FLXA analyzer using the WU11 interconnection cable for SENCOM™ sensors. The connections of this M9 male connector (see Section 2.7 for details), the WU11 interconnection cable and the FLXA terminal strip are given in Table 3.
6.2 Calibration for ORP and rH measurement For calibration of ORP and rH, the procedure for MANUAL CALIBRATION can be used as described in the Instruction Manual of the FLXA analyzer. The rH value is a function of the reference system and the pH value of the buffer solution. The FU24F sensor has a reference system of saturated Silver/Silver Chloride (Ag/AgCl).
7. MODEL CODE Model Code Suffix Code Option FU24F Model - NPT - FSM Description SENCOM™ pH Wide Body Sensor Dome shape model Flat surface model 8. SPARE PARTS Spare part K1521JA K1521JB Description FU24F SS holder 1” NPT PVDF holder 1” NPT K1520BA K1520BB K1520BC K1520BD K1520BE Buffer solutions Buffer solutions pH 4.01, pH 6.87, pH 9.18 (500 ml each) Buffer solution pH 1.68 (500 ml) Buffer solution pH 4.01 (500 ml) Buffer solution pH 6.87 (500 ml) Buffer solution pH 9.
9. EU DECLARATION OF CONFORMITY YOKOGAWA EU DECLARATION OF CONFORMITY Yokogawa Europe B.V.
Annex-1 Model Code Suffix Code Option Code Description FU24F SENCOM™ pH Wide body Sensor Model - NPT Dome shape model - FSM Flat surface model IM 12B6J8-01E-E
10. CHEMICAL COMPATIBILITY CHART Material Material PTFE PP PPS PVDF S.S.
YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us YOKOGAWA CHINA CO. LTD. 3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.