(BG) Всички улътвания за продукти от серията ATEX Ex ce лредлагат на английски език. Ако се нуждаете oт улътвания за продукти oт серията Ex на родния ви eзик, ce свържете c най-близкия oфис или представителство на фирма Yokogawa. (CZ) Všechny uživatelské příručky pro výrobky, na něž se vztahuje nevýbušné schváleni ATEX Ex, jsou dostupné v angličtině. Požadujete-li pokyny týkající se výrobků s nevýbušným schválením ve vašem lokálním jazyku, kontaktujte prosím vaši nejbližší reprezentačni kancelář Yokogawa.
(LV) Visas ATEX Ex kategorijas izstrâdâjumu Lietoðanas instrukcijas tiek piegâdâtas angïu valodâs. Ja vçlaties saòemt Ex ierîèu dokumentâciju citâ valodâ, Jums ir jâsazinâs ar firmas Jokogava (Yokogawa) tuvâko ofisu vai pârstâvi. (LT) Visos gaminiø ATEX Ex kategorijos Eksploatavimo instrukcijos teikiami anglø kalbomis. Norëdami gauti priestaisø Ex dokumentacijà kitomis kalbomis susisiekite su artimiausiu bendrovës Yokogawa biuru arba atstovu.
IM 12B6J3-03E-E
CONTENTS 1. PREFACE 1.1 Introduction 1.2 Unpacking and Checking 1.3 Warranty and Service 1.4 Serial number 6 6 6 7 7 2. GENERAL SPECIFICATIONS 2.1 Measuring elements 2.2 Wetted parts 2.3 Functional specifications (at 25°C) 2.4 Dynamic specifications 2.5 Operating range 2.6 Regulatory standards 2.7 Shipping details 2.8 Environmental conditions 2.9 Mechanical specifications 8 8 8 8 8 8 9 9 9 9 3. INSTALLATION OF FU24 3.1 Typical installation 3.2 Preparing the sensor for use 3.
1. PREFACE 1.1 Introduction This instruction manual provides information for the installation and use of the FU24 four-in-one wide body pH sensor, which is especially designed for harsh process applications. The Model FU24 offers a simple and cost effective solution for process applications with fluctuating pressure and/or temperature which normally limit the lifetime of a sensor. By using the patented Bellow system a strong pressure compensation mechanism is created.
1.3 Warranty and Service Yokogawa products and parts are guaranteed free from defects in workmanship and material under normal use and service for a period of (typically) 12 months from the date of shipment from the manufacturer. Individual sales organizations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted.
2. GENERAL SPECIFICATIONS 2.1 Measuring elements : pH glass electrode : Silver chloride reference : Solid platinum electrode : Pt1000 temperature sensor. 2.2 Wetted parts Sensor body Earthing pin Measuring sensor LE glass tube O-ring Reference junction : : : : : : PPS 40GF (Ryton™ with glass filling) Solid Platinum G-glass AR-glass Viton Porous PTFE 2.
2.6 Regulatory standards CE - ATEX - Certificate no. - Electrical data - Special - conditions (X) - Pressure - Applying article IECEx Applying standards Certificate no. : Decision 768/2008/EC : Directive 94/9/EC, as amended by Regulation (EC) no. 1882/2003 :D EKRA 11ATEX0014 X II 1 G Ex ia IIC T3...T6 Ga : For sensor input circuit connected to a certified intrinsically safe circuit with the following maximum values: Ui = 18 V; Ii = 170 mA; Pi = 0.4 W; Li = 0 mH; Ci = 0 nF (VP type) or 3.
3. INSTALLATION OF FU24 For optimum measurement results, the FU24 should be installed in a location that offers an acceptable representation of the process composition and DOES NOT exceed the specifications of the sensor. The FU24 is designed with 1” NPT threaded connections on both ends of the sensor to allow installation in a wide variety of applications. 3.1 Typical installation The FU24 sensor is designed for versatile in-line, immersion or off-line installation.
PG7 FU24 TeflonTape Figure 2. FF20-.
4. DIMENSIONS Dimensions in mm (inches) Figure 3.
5. WIRING 5.1 Connections The FU24 sensor is provided with a fixed cable or Variopin (VP) connector. The connections of both the wire as the VP connector are given in Figure 4. FU24 11/red, (Temp-1) 12/blue, (Temp-2) 13/black, (Ref.) ø6.5 15/clear, (pH) MODEL CODE Serie nummer 16/brown, (pH shield) 14/yellow, (ORP/LE) WU10-V VP-connector (Temp-1) Red (pH) A (Ref.) B (pH shield) C D (ORP/LE) F (Temp-2) E (Temp-1) (Temp-2 ) Blue (pH shield) Black (pH) Clear ø6.5 MODEL CODE Serie nummer (Ref.
5.3 Settings for PH402G and PH202G(S) Function pH pH compensated pH&ORP pH&rH ORP Setting (default) ORP code 01 0(pH) 0(pH) 0(pH) 1(ORP) 1(ORP) code 02 0(off) 1(ORP) 2(rH) 0(off) 0(off) code 03 1.1.1 1.1.1 1.1.1 0.0.1 0.0.1 code 04 0.0.1 0.0.1 0.0.1 0.0.1 1.1.1 code 31 0 2 2 jumper input 1 no no no yes yes jumper input 2 yes yes yes yes no 5.
6.2 Calibration for ORP and rH measurement For calibration of ORP and rH, the procedure for MANUAL CALIBRATION can be used as described in the Instruction Manual of the analyzer. The rH value is a function of the reference system and the pH value of the buffer solution. The FU24 sensor has a reference system of saturated Silver/Silver Chloride (Ag/AgCl). The commonly used standards for ORP and rH calibration are made from Chinhydron (Quinhydrone) powder dissolved in pH buffer solutions (1 g / 200 ml).
7. MODEL CODES Model Code Suffix Code Option FU24 Type - VP - 05 - 10 Temp. element - T1 Model - NPT - FSM Description Wide body sensor Variopin connector Fixed cable, 5 meter Fixed cable, 10 meter Pt1000 Dome shape model Flat surface model 8. SPARE PARTS Spare part Description FU24 K1521JA SS holder 1” NPT K1521JB PVDF holder 1” NPT Buffer solutions K1520BA Starters Kit: (3x 500 ml) Solution pH 4.01, pH 6.87, pH 9.18 K1520BB Buffer Solution (500 ml) pH 1.
9. EU DECLARATION OF CONFORMITY YOKOGAWA EU DECLARATION OF CONFORMITY Yokogawa Europe B.V.
Annex-1 Model Code Suffix Code Option FU24 Type - VP - 05 - 10 Temp.
10. CHEMICAL COMPATIBILITY CHART PFA 20 60 20 60 100 20 60 100 XXX OOO XXX OOO XXX OOO - - - OOO - - - OOO - - - XXX OOX XXX - - OOO - - OOO - - - - - OOO XXX OOO OOO XX - - - OOO - - - OOO OOX - - OOX - - XXX - - OOO OOO OOO OOO OOO OOO OOO XXX OOO XXX XXX OOO XXX OOO - - - OOO OOO - - OOO OOO XXX OOO OOO OOO XXX OOO XXX OOO XXX OOO - - - OOO OOO OOO XXX OOX OOO OOO X X X OOO X - OOO OOO OOO - - OOX XXX OOO OOO PVC 20 60 10 OOO 50 OOO 95 OX fuming - - Hydrochloric acid 10 OOO sat.
YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us YOKOGAWA CHINA CO. LTD. 3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.