Instruction Manual Model FU20F pH/ORP SENCOM™ sensor IM 12B6J3-04E-E 3rd edition
(BG) Всички улътвания за продукти от серията ATEX Ex ce лредлагат на английски език. Ако се нуждаете oт улътвания за продукти oт серията Ex на родния ви eзик, ce свържете c най-близкия oфис или представителство на фирма Yokogawa. (CZ) Všechny uživatelské příručky pro výrobky, na něž se vztahuje nevýbušné schváleni ATEX Ex, jsou dostupné v angličtině. Požadujete-li pokyny týkající se výrobků s nevýbušným schválením ve vašem lokálním jazyku, kontaktujte prosím vaši nejbližší reprezentačni kancelář Yokogawa.
(LV) Visas ATEX Ex kategorijas izstrâdâjumu Lietoðanas instrukcijas tiek piegâdâtas angïu valodâs. Ja vçlaties saòemt Ex ierîèu dokumentâciju citâ valodâ, Jums ir jâsazinâs ar firmas Jokogava (Yokogawa) tuvâko ofisu vai pârstâvi. (LT) Visos gaminiø ATEX Ex kategorijos Eksploatavimo instrukcijos teikiami anglø kalbomis. Norëdami gauti priestaisø Ex dokumentacijà kitomis kalbomis susisiekite su artimiausiu bendrovës Yokogawa biuru arba atstovu.
IM 12B6J3-04E-E
CONTENTS 1. 2. PREFACE 6 1.1 1.2 1.3 1.4 6 6 6 7 GENERAL SPECIFICATIONS 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10 3.
1. PREFACE 1.1 Introduction This instruction manual provides information for the installation and use of the FU20F SENCOM™ sensor. This digital sensor shows how Yokogawa applies the motto “Simply the Best” to sensor technology. Setup of the sensor is very easy because all sensor specific characteristics, such as calibration data, are stored in the sensor.
•C omplete shipping and billing instructions for return of material, plus the name and phone number of a contact person that can be reached for further information. • Clean Statement Returned goods that have been in contact with process fluids must be decontaminated and disinfected prior to shipment. Goods should carry a certificate to this effect, for the health and safety of our employees.
2. GENERAL SPECIFICATIONS 2.1 Measuring elements pH glass electrode Silver Chloride reference Solid Platinum electrode Pt1000 temperature sensor 2.2 Wetted parts Sensor body Measuring sensor Reference junction Earth pin O-ring : : : : : PPS 40GF (RytonTM with glass filling) G-glass Porous PTFE Solid Platinum Viton 2.
2.5 Operating range pH : 0 to 14 ORP : -1500 to 1500 mV rH : 0 to 100 Temperature - Dome shape : -10 °C to 105 °C (14 °F to 221 °F) - Flat surface : +15°C to 105°C (59 °F to 221 °F) Pressure : p(bar) 10 5 0 -10 Conductivity 0 25 50 105 Temp. (ºC) : > 50 µS/cm Note: T he pH operating range at room temperature is 0-14 pH, but at high temperatures or range outside 2-12 pH the lifetime will be seriously shortened. 2.
2.8 Regulatory standards CE - ATEX - Certificate no. - Electrical data - Special - conditions (X) : Decision 768/2008/EC : Directive 94/9/EC, as amended by Regulation (EC) no. 1882/2003 : DEKRA 11ATEX0064 X II 1 G Ex ia IIC T3...T6 Ga : For sensor input circuits (by connector) connected to a certified intrinsically safe circuit with the following maximum values : Ui = 6.1 V; Ii = 230 mA; Pi = 1.
CSA Certificate no. Master contract no. : 2516979 : 182892 IS, Class I Div. 1, GP A, B, C, D T3…T6 Electrical data : For sensor input circuits (by connector), connected to a certified intrinsically safe circuit, with the following maximum values : Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF or Certified intrinsically safe Yokogawa transmitter Model FLXA21 series. Ambient temperature : T6 for Tamb. -40 °C to +60 °C T5 for Tamb. -40 °C to +75 °C T4 for Tamb.
FM Certificate no. :3 046277 IS, Class I Div. 1, GP A, B, C, D T3…T6 Electrical data : For sensor input circuits (by connector), connected to a FM approved intrinsically safe apparatus meeting the entity parameters of the SENCOM sensor: Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF or FM approved intrinsically safe Yokogawa transmitter Model FLXA21 series. Ambient temperature: T6 for Tamb. -40 °C to +60 °C T5 for Tamb. -40 °C to +75 °C T4 for Tamb. -40 °C to +85 °C T3 for Tamb.
Note: When the sensor has been connected to none intrinsically safe equipment which exceeds the restrictions regarding the sensor input circuit (see electrical data), the sensor is not suitable anymore for intrinsically safe use. 2.9 Shipping details Package size (L x W x H) : 300 x 100 x 75 mm (11.8 x 3.9 x 3.0 inch) Package weight : app. 0.33 kg (0.73 lbs) 2.
3. INSTALLATION OF FU20F For optimum measurement results, the FU20F should be installed in a location that offers an acceptable representation of the process composition and DOES NOT exceed the specifications of the sensor. The FU20F is designed with 3/4” NPT threaded connections on both ends of the sensor to allow installation in a wide variety of applications. 3.1 Typical installation The FU20F sensor is designed for versatile in-line, immersion or off-line installation.
The FU20F can also be mounted using one of the optional quick-removal adapters (/NSS, /NTI, /BSS or /BTI). For a detailed description of these adapters see Sections 4 and 7 of this Instruction Manual. 1 Apply Teflon tape to the threaded end of this adapter; 2 I nstall the adapter in the process connection. Tighten the adapter using a wrench on the adapter flats.
FD40 Fitting Cleaning Solution (water, acid,etc.
4. DIMENSIONS Dimensions in mm (inches) 3/4” TAPERED THREAD 3/4” TAPERED THREAD Ø 26.0 (1.02”) -NPT 19.6 (0.78”) 19.6 (0.78”) 22.5 (0.89”) 1” NPT or ISO 7/1-R1 /NSS, /NTI, /BSS, /BTI 51.0 (2.1”) -FSM 21 (0.83”) Ø 22.0 (0.86”) 150.0 (5.9”) 65.5 (2.58”) 10.0 (0.39”) Figure 9: Dimensions of FU20F Sensor Ø 35.5 (1.4”) 25.0 (1”) 99.0 (3.9”) /NSS, /NTI, /BSS, /BTI 1” NPT or ISO 7/1-R1 Ø 39.0 (1.
5. WIRING The FU20F is provided with a multipole M9 male connector for connection to the Yokogawa FLXA analyzer using the WU11 interconnection cable for SENCOM™ sensor. The connections of this M9 male connector (see Section 2.7 for details), the WU11 interconnection cable and the FLXA terminal strip are given in Table 3.
6.2 Calibration for ORP and rH measurement For calibration of ORP and rH, the procedure for MANUAL CALIBRATION can be used as described in the Instruction Manual of the FLXA analyzer. The rH value is a function of the reference system and the pH value of the buffer solution. The FU20F sensor has a reference system of saturated Silver/Silver Chloride (Ag/AgCl).
7.
9. EU DECLARATION OF CONFORMITY EU EUDECLARATION DECLARATIONOF OFCONFORMITY CONFORMITY We: Europe B.V. We: Yokogawa Yokogawa Europe B.V.
Annex-1 Model Code Suffix Code Option Code Description FU20F SENCOM™ pH Wide body Sensor Model - NPT Dome shape model - FSM Flat surface model Options /HCNF Hastelloy cleaning system /FPS Adapter for FF40, FS40, FD40 fittings, Noryl /NSS 1” NPT adapter, Stainless Steel (316L) /NTI 1” NPT adapter, Titanium /BSS 1” BSP adapter, Stainless Steel (316L) /BTI 1” BSP adapter, Titanium IM 12B6J3-04E-E
10. CHEMICAL COMPATIBILITY CHART 20 60 20 60 100 20 60 100 XXX OOO XXX OOO XXX OOO - - - OOO - - - OOO - - - XXX OOX XXX - - OOO - - OOO - - - - - OOO XXX OOO OOO XX - - - OOO - - - OOO OOX - - OOX - - XXX - - OOO OOO OOO OOO OOO OOO OOO XXX OOO XXX XXX OOO XXX OOO - - - OOO OOO - - OOO OOO XXX OOO OOO OOO XXX OOO XXX OOO XXX OOO - - - OOO OOO OOO XXX OOX OOO OOO X X X OOO X - OOO OOO OOO - - OOX XXX OOO OOO PFA 20 60 10 OOO 50 OOO 95 OX fuming - - Hydrochloric acid 10 OOO sat.
YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us YOKOGAWA CHINA CO. LTD. 3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.