Owner manual
IM 12B6J3-02E-E
7
4. WIRING
(pH) A
(Ref.)
B
(pH shield) C
(ORP/LE)
D
(Temp)
E
(Temp)
F
VP-connector
4.1. Cable marking and pin allocation.
FU20 cable
Wire color
WU10-V
cable
Wire color
Signal description
Cable wires
(&EXA terminal)
number
VP connector
Pin
pH Clear 15 A Clear
pH-Shield Brown 16 C Black
Reference Black 13 B Brown
ORP/Liquid earth Yellow 14 D Yellow
Temp-1 Red 11 E Red
Temp-2 Blue 12 F Blue
4.2. Instrument connection of sensor as function of measurement
Function pH pH&ORP pH&rH ORP pH compen-
Signal descr. sated ORP
pH 15 15 15 - 13
Shield of pH 16 16 16 - 17
REF 13 13 13 13 -
ORP/Liquid Earth 14 14 14 15 15
TEMP 11 11 11 11 11
TEMP(2) 12 12 12 12 12
Jumper - - - 13-14 14-15
The settings for the pH converters must be set according to the instructions in the
Instruction Manual of the converter. In the following table these settings are summarized.
(pH)
Clear A
(Ref.) Brown B
(ORP/LE) Yellow D
(Temp-1) Red
F
(Temp-2 ) Blue
E
(pH shield) Black C
FU20
WU10
MODEL CODE
Serie nummer
Red (11), Temp-1
ø6.5
Blue (12), Temp-2
Black (13), Reference
Clear (15), pH
Brown (16), pH shield
Yellow (14), ORP/LE
MODEL CODE
Serie nummer
ø6.5