Operation Manual
<5. Operational Flowchart>
5-7
IM 01C25R51-01E
(A) New Conguration
Step FieldMate Menu
DYF(SoftDL) F
OUNDATION fieldbus DTM
Layer1 Layer2 Layer3
DYFMVTool
Main Window
Procedure
"Online Parameter" is opened.
Upload "Device Flow Parameters"
and "Device General Parameters".
Edit "Device General Parameters"
and Apply "Offline Database". 3)
"Initialize Menu" (If needed)
Initialize "User Flow Parameters"
and "Device Flow Parameters".
Select compensation and
edit "Device General Parameter".
Edit "User Flow Parameters".
Edit "User Flow Parameters".
Edit "User Flow Parameters" and
"Device Flow Parameters".
"Export -> Export VMV Menu"
Export "User Flow Parameters"
and "Device Flow Parameters" to
binary file.
Download
"Device Flow Parameters" and
"Device General Parameters".
Confirm the compensation
calculation by input simulation for
flow rate, temperature and
pressure.
"Report"
Export "User Flow Parameters" in
CSV file format.
Apply "Device Flow Parameters"
to "Offline Database".
Goes offline.
Save the "*.dns"
"User Flow Parameters", "Device
Flow Parameters" and "Device
General Parameters".
1
Start DTM and
upload
parameter
from the
device
Perform Flow
Configuration
Download
parameter to
the device
Confirm flow
configuration
Save settings
to DNS file
2
3
4
5
Device
Device
Parameter
Offline
Parameter
Flow
Configuration
Wizard
Flow
Configuration
Mode
Fluid Type
Setup
Fluid
Operating
Range Setup
Fluid Physical
Property
Setup
Apply Flow
Configuration
Assigned DTM
Device
Download to
device 2)
Upload from
device 1)
Parameter
Additional
Functions
Online
Parameter
Disconnect
Save to file...
Device
File
F0505E.ai
1) Beforestartingoperation,execute‘Uploadfromdevice’,sothat“DeviceGeneralParameters”oftheowmeterin
theofinedatabasearesynchronizedwiththedevice.Uploaddatafromthedevicedoesnotinclude“UserFlow
Parameters”.
*:UserFlowParametersconsistofuidinformationandoperatingcondition.
2) Executing‘Downloadtodevice’willwriteboth“DeviceFlowParameters”and“DeviceGeneralParameters”into
thedevice.Beforeexecuting‘Downloadtodevice’,execute‘Uploadfromdevice’,soastomake“DeviceGeneral
Parameters”oftheowmeterintheofinedatabasesynchronizedwiththedevice.
Downloaddatatothedevicedoesnotinclude“UserFlowParameters”.
3) TemperatureunitandpressureunitinARBlockarenotconguredinFlowCongurtionWizardDetailCompensation
Mode.
TheseparametersmustbesetmanuallyinARBlockofineparameter.