Step 5 Conrm input conditions in AR Block
Conrmthatthefollowingparametersarethesamevaluesinputin<Step4>.
“Input.Value”,“Input1.Value”and“Input2.Value”
([OnlineParameter]→[Arithmetic]→[DynamicVariables])
Refer to Table 8.10.
IMPORTANT
Set"Good"statusinthefollowingparametersbeforeinputtingvalues:
“Input.Status”,“Input1.Status”and“Input2.Status”
([OnlineParameter]→[Arithmetic]→[Conguration]→[InputParameters])
IMPORTANT
In DYF(SoftDL) FOUNDATIONeldbusDTM,"Input.Value","Input1.Value"and"Input2.Value"are
denedasnon-volatileparameters.
In Device DTM, the update time for non-volatile parameters on the Online Parameter screen is
s
etto300secondsasdefault.Setthevalueof"RefreshtimeforSTATICs"toabout30seconds
becauselongertimeisneededtodisplayparametersinthedefaultconguration.
(DeviceDTMmenu:[Device]→[Additionalfunctions]→[Communication])
Step 6 Conrm output in AR Block
Refer to section 8.3.1 <Step 4>