Operation Manual
<8. Operation of DYFMVTool>
8-21
IM 01C25R51-01E
IMPORTANT
Beforeinputtingvalues,set"Good_NonCascade::NonSpecic:NotLimited"statusinthefollowing
parameters:
“
Input.Status”,“Input1.Status”and“Input2.Status”
([OnlineParameter]→[Arithmetic]→[Conguration]→[InputParameters])
*: Toshow"Good_NonCascade::NonSpecic:NotLimited",openthepull-downmenuofeach
Statusparameterandpressthe[G]key.
Step 4 Conrm output in AR Block
Conrmthevalueofthefollowingoutputparameters(RefertoFigure8.19andTable8.11):
“DensityFactor.Value”and”PreOutput.Value”.
([OnlineParameter]→[Arithmetic]→[DynamicVariables])
Table 8.11 Output in AR Block
Output Device Function Block
Parameter
Value Unit
Density
digitalYEWFLO
AR Block
(Computing Section)
DENSITY_FACTOR
D
ensityFactor.Value
(Relative Index: 48)
<Effective Mode: Auto/MAN/OOS>
Density Factor Unit
(Relative Index: 49)
Mass Flow
Rate
AR Block
(Output Section)
PRE_OUT
P
reOutput.Value
(Relative Index: 9)
<Effective Mode: Auto/MAN/OOS>
Output Range.Units
Index
(Relative Index: 11)
IMPORTANT
Make sure that no alarming status of AR Block is shown in Device Status, and then confirm the
value of output parameters. Refer to Appendix B-3 for alarm.
(
[OnlineParameter]→[DeviceStatus]→[DiagnosticList])
IMPORTANT
Makesure"Good_NonCascade::NonSpecific:NotLimited"statusofthefollowingparameters
beforeconrmingthevalues:
“DensityFactor.Status”and“PreOutput.Status”
([OnlineParameter]→[Arithmetic]→[DynamicVariables])
NOTE
Conrmthesettingsin"DensityFactorUnit"and"OutputRange.UnitIndex".
([OnlineParameter]→[Arithmetic]→[Conguration]→[DensityFactorParameters])
([OnlineParameter]→[Arithmetic]→[Conguration]→[OutputParameters])