1) For details of the calibration procedures, follow User’s Manual FY2-XJAM012-40-2E.
Record the following original parameters and set the AXF parameters as shown below
manually. (These settings are described in FY2-XJAM012-40-2E.)
Table 6.3.10 AXF parameter setting of Converter Check
HART Parameter (DD)
Parameter of
AXF Display
Setting Value
Select Flow
Tube (*1)
Device setup Detailed Setup Basic SetupSe-
lect Flow Tube
C30 Calibrator
Nominal Size
Device setup Detailed Setup Basic Setup
→Nominal Size
C32
Enter owtube
nominal size in
“mm” unit
Nominal Size
Device setup Detailed Setup Basic Setup
→Nominal Size unit
C31 mm
MF Set
Device setup Detailed Setup Basic Setup→MF
Set
C20
(Measure Mode)
Standard DF
or
Enhanced DF (*3)
Low MF
Device setup Detailed Setup Basic Setup→MF
Set Standard DF Low MF
C21 1.0
High MF
Device setup Detailed Setup Basic Setup→MF
Set Standard DF High MF
C22 1.0
Low MF (EDF)
(*2)
Device setup Detailed Setup Basic Setup→MF
Set Enhanced DF Low MF (EDF)
C23 1.0
High MF (EDF)
(*2)
Device setup Detailed Setup Basic Setup→MF
Set Enhanced DF High MF (EDF)
C24 1.0
*1: Only for AXFA11
*2: Only for optional code HF1 or HF2 (Enhanced Dual Frequency Excitation)
*3: Only for optional code HF1 or HF2, select “Enhanced DF”.
2) Enter the span on the Converter” screen.
2)-1 When Verication Tool is in “onlinemode:
Click “Read Span (3)in Figure 6.3.15. The value of parameter “Velocity Check” is entered
automatically. If the AXF parameters are different from those in Table 6.3.10, the following
error message is displayed. Set the AXF parameters to those in Table 6.3.10.
F060317.ai
Figure 6.3.17 Example of “Parameter Error” Dialog