Manual

<6 OPERATION>
6-32
IM 01R01A11-01EN
(B) Active Pulse Output (with optional code EM)
1) Wire AXF and CA150 Handy Calibrator or similar as shown in Figure 6.3.13.
2) Select “2 pps” in “Setting Value”. See Figure 6.3.11.
F060311.ai
Figure 6.3.11 “Setting Value” in “Active Pulse Output”
3) Output
3)-1 When Veri cation Tool is in “online” mode:
Select “Pulse” in “Output Selection” and click “Output”. See Figure 6.3.12.
F060312.ai
Figure 6.3.12 “Output Selectionin “Active Pulse Output”
3)-2 When Veri cation Tool is in “of inemode:
3)-2-1 Record the original AXF parameters, and then set them as shown below.
Table 6.3.7 Parameter Settings of “Active Pulse Output”
HART Parameter (DD)
Parameter of
AXF Display
Setting Value
Pulse Scale Device Setup Easy Setup Pulse Scale B33/E11 2
Pulse Width Device Setup Pulse Set Pulse Width E12 100ms
Pulse Unit Device Setup Easy Setup Pulse Scale B32/E10 Pulse/s
Test DO (*1) Device Setup Diag/Service Output Test→ Test DO N30 Pulse
DO Function(*1)
Device Setup Con guration Status Function
DO Function
F20 Pulse Output
(*1) only for AXFA14 or AXF Integral Flowmeter
3)-2-2 Output pulse by setting AXF HART parameter “Loop testto 20 mA”.
Table 6.3.8 Parameter Setting of “Active Pulse Output
HART Parameter (DD)
Parameter of
AXF Display
Setting Value
Loop test Device Setup Diag/Service Output TestLoop test - 20 mA