Manual

<6 OPERATION>
6-30
IM 01R01A11-01EN
NOTE
Measure for more than 60 seconds.
If using a YOKOGAWA CA150, set the mode to “Contact Input” and the range to “CPM”.
4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or
equivalent to simplify the measurement of pulses. If such a device is unavailable, an appro-
priate load (30V DC, 0.2A max) is required and use the con guration shown in Figure 6.3.9.
SBBASACAL–AL+COMSO2+SO1+I+ I–
CUR OUT STATUS OUT ALARM OUT SIGNAL
COMSI2+SI1+EX2EX1L/+N/–
POWER SUPPLY EXCITER
P–P+
PULSE OUT STATUS IN
Electronic Counter
30VDC , 0.2A. max
External
power supply
Load
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
F060309-A.ai
(A) AXFA11
Electronic Counter
30VDC , 0.2A. max
External
power supply
Load
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
F060309-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.9 Pulse Output Diagram (without optional code EM)
5) Enter the measured value manually.
6) “Result” is displayed automatically.
F060310.ai
Figure 6.3.10 “Pulse OutputResult