<6 OPERATION>
6-26
IM 01R01A11-01EN
Current Output
Use a Yokogawa CA150 Handy Calibrator or similar.
Set the Digital MultiMeter (CA150 or similar) to measure the current in “mA”.
NOTE
SBBASACAL–AL+COMSO2+SO1+
I+ I–
CUR OUT STATUS OUT ALARM OUT SIGNAL
COMSI2+SI1+EX2EX1L/+N/–
POWER SUPPLY EXCITER
P–P+
PULSE OUT STATUS IN
250 to 600
Digital
Multi Meter
F060303-A.ai
(A) AXFA11
250 to 600
Digital
Multi Meter
Current Output
Use a Yokogawa CA150 Handy Calibrator or similar.
Set the Digital MultiMeter (CA150 or similar) to measure the current in “mA”.
NOTE
F060303-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.3 Current Output Connection Diagram