6.2.2 Physical Appearance
2
3
1
F060211.ai
Figure 6.2.11 “Physical Appearance” Screen
Table 6.2.9 “Physical Appearance” Screen
Item
No.
Item Name Description
1 Note Opens another window as shown in Figure 6.2.12. Click “Note”.
2 Flowtube Selects the result of the Flowtube Appearance Check.
3 Converter Selects the result of the Converter Appearance Check.
Physical Appearance
A physical observation of the Flowmeter is required, check the following items:
Flowtube
Check the Flowtube and the Cables connected to the Flowtube for:
* Deformation, Damage, Corrosion, Cracking, Signs of Abrasion and any Leaks.
Converter
Check the Converter and the Cables connected to the Converter for:
* Corrosion, Damage, moisture in the housing, corroded terminals.
Observations are objective.
F060212.ai
Figure 6.2.12 Example of Physical AppearanceNote
1) Check the physical appearance of “Flowtubeand “Converter”.
For details, see Figure 6.2.12.
2) Select an applicable result manually in Table 6.2.10.
Table 6.2.10 Physical Appearance
Result Description
Passed Is selected when the result is good and there is no problem.
Failed Is selected when the result is bad.
N/A Is selected when the appearance check is not possible.
3) “Total Resultis displayed automatically. See Figure 6.2.13.