<6 OPERATION>
6-19
IM 01R01A11-01EN
1
2
F060208.ai
Figure 6.2.8 “Flow Conrmation” Dialog
Table 6.2.7 “Flow ConrmationDialog
Item
No.
Item Name Description
1 OK Runs “Circuitcheck and “Device Status” check.
2 Cancel
Stops runningCircuit” check and “Device Status” check, and displays
“Main Screen”.
4) Click OK (1)” after conrming that the owtube is lled with uid and the uid velocity is suf-
cient. “Circuit” check andDevice Status” check are run. During these checks, the status
bar and remaining time are displayed. See Figure 6.2.9. Regarding the execution time, see
Table 6.2.8.
F060209.ai
Figure 6.2.9 “During ”Circuit” Check and “Device StatusCheck” Screen
Table 6.2.8 Execution time [Flow]
No. Optional code Execution time
1 Without ELC 630 s
2 With ELC (DC Noise Cut Circuit) 120 s